Anode sensitivity calibration of photomultiplier tubes used as a readout of the ELOSS detector
We present and discuss the methodology and results for an anode sensitivity calibration of Photomultiplier Tubes (PMTs) that serve as the optical readout of the Energy Loss Optical Scintillation System (ELOSS). The ELOSS detector, under development at the Facility for Rare Isotope Beams (FRIB), is a gaseous Xe scintillation-based detector designed to identify the atomic number (Z) of nuclear reaction fragments by energy-loss measurement. Variation in the anode sensitivity from tube to tube of commercially available PMTs limits ELOSS from reaching its ultimate Z resolution. A methodical approach to calibrating PMT anode sensitivities to a single target value by systematic PMT gain corrections is presented. Although the calibration process is discussed in terms of its impact on ELOSS performance, the presented method is applicable to any optical detector system composed of a large number of PMTs needing anode sensitivity calibration.