In-Flight Calibration of Hitomi Soft X-Ray Spectrometer (3) Effective Area

We present the result of the in-flight calibration of the effective area of the Soft X-ray Spectrometer (SXS) on board the Hitomi X-ray satellite using an observation of the Crab nebula. We corrected for artifacts when observing high count rate sources with the X-ray microcalorimeter. We then constructed a spectrum in the 0.5–20 keV band, which we modeled with a single power-law continuum attenuated by interstellar extinction. We evaluated the systematic uncertainty of the spectral parameters by various calibration items. In the 2–12 keV band, the SXS result is consistent with the literature values in flux (2.20 ± 0.08 × 10(exp −8) erg s(exp-1) cm(exp −2) with a 1 σ statistical uncertainty) but is softer in the power-law index (2.19 ± 0.11). The discrepancy is attributable to the systematic uncertainty of about +6%/−7% and +2%/−5% respectively for the flux and the power-law index. The softer spectrum is affected primarily by the systematic uncertainty of the Dewar gate valve transmission and the event screening.

Tsujimoto, Masahiro↗

The XRS Low Temperature Cryogenic System: Ground Performance Test Results

The X-Ray Spectrometer (XRS) instrument is part of the Astro-E mission scheduled to launch early in 2000. Its cryogenic system is required to cool a 32-element square array of x-ray microcalorimeters to 60-65 mK over a mission lifetime of at least 2 years. This is accomplished using an adiabatic demagnetization refrigerator (ADR) contained within a two-stage superfluid helium/solid neon cooler. Goddard Space Flight Center is providing the ADR and helium dewar. The flight system was assembled in Sept. 1997 and subjected to extensive thermal performance tests. This paper presents test results at both the system and component levels. In addition, results of the low temperature topoff performed in Japan with the engineering unit neon and helium dewars are discussed.

Breon, Susan↗

Multiabsorber Transition-Edge Sensors for X-Ray Astronomy

We are developing arrays of position-sensitive microcalorimeters for future x-ray astronomy applications. These position-sensitive devices commonly referred to as hydras consist of multiple x-ray absorbers, each with a different thermal coupling to a single-transition-edge sensor microcalorimeter. Their development is motivated by a desire to achieve very large pixel arrays with some modest compromise in performance. We report on the design, optimization, and first results from devices with small pitch pixels (<75 μm) being developed for a high-angular and energy resolution imaging spectrometer for Lynx. The Lynx x-ray space telescope is a flagship mission concept under study for the National Academy of Science 2020 decadal survey. Broadband full-width-half-maximum (FWHM) resolution measurements on a 9-pixel hydra have demonstrated ΔE(FWHM) = 2.23 ± 0.14 eV at Al-Kα, ΔE(FWHM) = 2.44 ± 0.29 eV at Mn-Kα, and ΔE(FWHM) = 3.39 ± 0.23 eV at Cu-Kα. Position discrimination is demonstrated to energies below <1 keV and the device performance is well-described by a finite-element model. Results from a prototype 20-pixel hydra with absorbers on a 50-μm pitch have shown ΔE(FWHM) = 3.38 ± 0.20 eV at Cr-Kα1. We are now optimizing designs specifically for Lynx and extending the number of absorbers up to 25/hydra. Numerical simulation suggests optimized designs could achieve ∼3 eV while being compatible with the bandwidth requirements of the state-of-the art multiplexed readout schemes, thus making a 100,000 pixel microcalorimeter instrument a realistic goal.

Smith, Stephen J.↗

In-Flight Performance of the Soft X-Ray Spectrometer Detector System on ASTRO-H

The SXS instrument was launched aboard the Astro-H observatory on February 17, 2016. The SXS spectrometer is based on a high sensitivity x-ray calorimeter detector system that has been successfully deployed in many ground and sub-orbital spectrometers. The instrument was to provide essential diagnostics for nearly every class of x-ray emitting objects from the atmosphere of Jupiter to the outskirts of galaxy clusters, without degradation for spatially extended objects. The SXS detector system consisted of a 36-pixel cryogenic microcalorimeter array operated at a heat sink temperature of 50 mK. In pre-flight testing, the detector system demonstrated a resolving power of better than 1300 at 6 keV with a simultaneous band-pass from below 0.3 keV to above 12 keV with a timing precision better than 100 microsecond. In addition, a solid-state anti-coincidence detector was placed directly behind the detector array for background suppression. The detector error budget included the measured interference from the SXS cooling system and the spacecraft. Additional margin for on-orbit gain-stability, and on-orbit spacecraft interference were also included predicting an on-orbit performance that meets or exceeds the 7 eV FWHM at 6 keV requirement. The actual on-orbit spectral resolution was better than 5 eV FWHM at 6 keV, easily satisfying the instrument requirement. Here we discuss the actual on-orbit performance of the SXS detector system and compare this to performance in pre-flight testing and the on-orbit predictions. We will also discuss the on-orbit gain stability, additional on-orbit interference, and measurements of the on-orbit background.

X-ray astrophysics↗

Operation of an ADR Using Helium Exchange Gas as a Substitute for a Failed Heat Switch

The Soft X-ray Spectrometer (SXS) is one of four instruments on the Japanese Astro-H mission, which is currently planned for launch in late 2015. The SXS will perform imaging spectroscopy in the soft X-ray band (0.3-12 keV) using a 6 6 pixel array of microcalorimeters cooled to 50 mK. The detectors are cooled by a 3-stage adiabatic demagnetization refrigerator (ADR) that rejects heat to either a superfluid helium tank (at 1.2 K) or to a 4.5 K Joule-Thomson (JT) cryocooler. Four gas-gap heat switches are used in the assembly to manage heat flow between the ADR stages and the heat sinks. The engineering model (EM) ADR was assembled and performance tested at NASA/GSFC in November 2011, and subsequently installed in the EM dewar at Sumitomo Heavy Industries, Japan. During the first cooldown in July 2012, a failure of the heat switch that linked the two colder stages of the ADR to the helium tank was observed. Operation of the ADR requires some mechanism for thermally linking the salt pills to the heat sink, and then thermally isolating them. With the failed heat switch unable to perform this function, an alternate plan was devised which used carefully controlled amounts of exchange gas in the dewar's guard vacuum to facilitate heat exchange. The process was successfully demonstrated in November 2012, allowing the ADR to cool the detectors to 50 mK for hold times in excess of 10 h. This paper describes the exchange-gas-assisted recycling process, and the strategies used to avoid helium contamination of the detectors at low temperature.

Adiabatic demagnetization↗

The Design, Implementation, and Performance of the Astro-H SXS Calorimeter Array and Anti-Coincidence Detector

The calorimeter array of the JAXA Astro-H (renamed Hitomi) Soft X-ray Spectrometer (SXS) was designed to provide unprecedented spectral resolution of spatially extended cosmic x-ray sources and of all cosmic x-ray sources in the Fe-K band around 6 keV, enabling essential plasma diagnostics. The SXS has a square array of 36 microcalorimeters at the focal plane. These calorimeters consist of ion-implanted silicon thermistors and HgTe thermalizing x-ray absorbers. These devices have demonstrated a resolution of better than 4.5 eV at 6 keV when operated at a heat-sink temperature of 50 mK. We will discuss the basic physical parameters of this array, including the array layout, thermal conductance of the link to the heat sink, resistance function, absorber details, and means of attaching the absorber to the thermistor-bearing element. We will also present the thermal characterization of the whole array, including thermal conductance and crosstalk measurements and the results of pulsing the frame temperature via alpha particles, heat pulses, and the environmental background. A silicon ionization detector is located behind the calorimeter array and serves to reject events due to cosmic rays. We will briefly describe this anti-coincidence detector and its performance.

x-ray calorimeter↗

High resolution X-ray spectroscopy using microcalorimeters

The use of microcalorimeters for high-resolution, high quantum efficiency, nondispersive X-ray spectroscopy has been demonstrated over the past few years. In this paper, the principles of X-ray calorimetry are reviewed, and the results of ongoing X-ray tests using microcalorimetry are summarized. An approach to building an X-ray calorimeter spectrometer is discussed.

Kelley, R. L.↗

Development of Transition Edge Sensor Detectors Optimized for Single-Photon Spectroscopy in the Optical and Near-Infrared

The search for biosignatures in the atmospheres of exoplanets will be a key focus of future space telescopes that operate in the ultraviolet, visible, and near-infrared bands. Detection of biosignatures requires an instrument with moderate spectral resolving power ($R \sim 100$) and a large bandwidth ($\sim 400$ nm -- $\sim 1.8$ $\mu$m). Additionally, biosignature detection is a photon-starved science; instruments designed for these measurements would ideally combine high optical efficiency with quantum-limited photon detectors (i.e., detectors that exhibit zero dark current). In this work, we report on our efforts to develop energy resolving transition edge sensor (TES)-based detectors designed for biosignature detection. TESs operated as microcalorimeters are compelling detectors for this application. Unlike semiconductor detectors, TESs eliminate the need for dispersive optics and are truly single photon detectors -- fundamental TES noise yields uncertainty in the energies of detected photons, not in the number of detected photons. We introduce TESs designed for this application and discuss the path toward realizing a TES-based dispersionless spectrometer optimized for biosignature detection.

Instrumentation and Photography↗

Development of Transition Edge Sensor Detectors Optimized for Single-Photon Spectroscopy in the Optical and Near-Infrared

The search for biosignatures in the atmospheres of exoplanets will be a key focus of future space telescopes that operate in the ultraviolet, visible, and near-infrared bands. Detection of biosignatures requires an instrument with moderate spectral resolving power (R ~ 100) and a large bandwidth (~ 400 nm – ~ 1.8 µm). Additionally, biosignature detection is a photon-starved science; instruments designed for these measurements would ideally combine high optical efficiency with quantum-limited photon detectors (i.e., detectors that exhibit zero dark current). In this work, we report on our efforts to develop energy resolving transition edge sensor (TES)-based detectors designed for biosignature detection. TESs operated as microcalorimeters are compelling detectors for this application. Unlike semiconductor detectors, TESs eliminate the need for dispersive optics and are truly single photon detectors – fundamental TES noise yields uncertainty in the energies of detected photons, not in the number of detected photons. We introduce TESs designed for this application and discuss the path toward realizing a TES-based dispersionless spectrometer optimized for biosignature detection.

Instrumentation and Photography↗

In-Orbit Operation of the ASTRO-H SXS

We summarize all of the in-orbit operations of the soft x-ray spectrometer (SXS) onboard the ASTROH (Hitomi) satellite. The satellite was launched on February 17, 2016, and the communication with the satellite ceased on March 26, 2016. The SXS was still in the commissioning phase, in which the set-ups were progressively changed. This paper is intended to serve as a concise reference of the events in orbit in order to properly interpret the SXS data taken during its short lifetime and as a test case for planning the in-orbit operation for future microcalorimeter missions.

Tsujimoto, Masahiro↗

In-orbit Operation of the Soft X-Ray Spectrometer Onboard the Hitomi Satellite

We summarize all of the in-orbit operations of the soft x-ray spectrometer (SXS) onboard the ASTRO-H (Hitomi) satellite. The satellite was launched on February 17, 2016, and the communication with the satellite ceased on March 26, 2016. The SXS was still in the commissioning phase, in which the set-ups were progressively changed. This paper is intended to serve as a concise reference of the events in orbit in order to properly interpret the SXS data taken during its short lifetime and as a test case for planning the in-orbit operation for future microcalorimeter missions.

ASTRO-H↗

Detector Assembly and the Ultralow-Temperature Refrigerator for XRS

The X-ray spectrometer (XRS) on the Japanese Astro-E Spacecraft is the first ultra low temperature space borne instrument. The system utilizes a 900g Ferric Ammonium Alum (FAA) Adiabatic Demagnetization Refrigerator (ADR) with a helium-3 gas gap heat switch to cool the detector assembly to 0.060K. The system operates in a "single shot" configuration allowing the system to remain at its operating temperature for about 40 hours in the lab. The on-orbit performance is expected to be about 35 hours with a 97% duty cycle. The detector assembly for XRS consists of a 32 channel microcalorimeter array bias electronics, thermometry, and an anti-coincidence detector that are attached to the cold stage of the ADR. To thermally Isolate the detector system from the superfluid helium reservoir, the detector system is suspended by Kevlar cords and electrical connection in made by L30, 17-micron diameter, tensioned NbTi leads. The detectors are read out in a source-follower arrangement using FET amplifiers operating at 130K mounted in multiply-thermally-isolated assemblies that also use Kevlar and stainless steel wiring. The design and thermal performance of this system will be discussed and compared to the theoretical limits.

Porter, F. S.↗

Laboratory Astrophysics Using a Spare XRS Microcalorimeter

The XRS instrument on Astro-E is a fully self-contained microcalorimeter x-ray instrument capable of acquiring optimally filtering, and characterizing events for 32 independent pixels. With the launch of the Astro-E spacecraft, a full flight spare detector system has been integrated into a laboratory cryostat for use on the electron beam ion trap (EBIT) at Lawrence Livermore National Laboratory. The detector system contains a microcalorimeter array with 32 instrumented pixels heat sunk to 60 mK using an adiabatic demagnetization refrio,erator. The instrument has a composite resolution of 8eV at 1 keV and 12eV at 6 keV with a minimum of 95% quantum efficiency. This will allow high spectral resolution, broadband observations of collisionally excited plasmas which are produced in the EBIT experiment. Unique to our instrument are exceptionally well characterized 1000 Angstrom thick aluminum on polyimide infrared blocking filters. The detailed transmission function including the edc,e fine structure of these filters has been measured in our laboratory using an erect field grating spectrometer. This will allow the instrument to perform the first broadband absolute flux measurements with the EBIT instrument. The instrument performance as well as the results of preliminary measurements will be discussed. Work performed under the auspices of the U.S. D.o.E. by Lawrence Livermore National Laboratory under contract W-7405-ENG-48 and was supported by the NASA High Energy Astrophysics Supporting Research and Technology Program.

Audley, M. Damian↗

The In-Flight Calibration Program for the XRS on Astro-E2

The X-Ray Spectrometer (XRS) will be launched in February 2005 as part of the Astro-E2 mission. It will provide unprecedented throughput and resolving powers particularly at high energies. In this presentation we will describe the in-flight calibration program. The energy scale of the XRS is a complex, non-linear function of the noise and power conditions on the array. It will be calibrated empirically using the bright point sources, Capella and GX301-2. Ground calibration of the line spread function show it to be almost perfectly Gaussian. The in-flight calibration is designed to verify this using the energy scale targets. The effective area curve of the XRS contains discreet edge structure from the mirrors, the optical blocking filters, and the microcalorimeter HgTe absorbers. The effective area Calibration program will simultaneously measure these absorption edges and the global effective area properties using the relatively featureless sources 3C273 and Mrk421. Additional monitoring of any ice build-up on the filters will be conducted using observations of the supernova remnants N132D and E0102.

Cottam, J.↗

Status of the Micro-X Sounding Rocket X-Ray Spectrometer

Micro-X is a sounding rocket borne X-ray telescope that utilizes transition edge sensors to perform imaging spectroscopy with a high level of energy resolution. Its 2.1m focal length X-ray optic has an effective area of 300 sq cm, a field of view of 11.8 arcmin, and a bandpass of 0.12.5 keV. The detector array has 128 pixels and an intrinsic energy resolution of 4.5 eV FWHM. The integration of the system has progressed with functional tests of the detectors and electronics complete, and performance characterization of the detectors is underway. We present an update of ongoing progress in preparation for the upcoming launch of the instrument.

X-ray astronomy↗

Automated Hybridization of X-ray Absorber Elements-A Path to Large Format Microcalorimeter Arrays

In the design of microcalorimeters, it is often desirable to produce the X-ray absorber separately from the detector element. In this case, the attachment of the absorber to the detector element with the required thermal and mechanical characteristics is a major challenge. In such arrays, the attachment has been done by hand. This process is not easily extended to the large format arrays required for future X- ray astronomy missions such as the New x-ray Telescope or NeXT. In this paper we present an automated process for attaching absorber tiles to the surface of a large-scale X-ray detector array. The absorbers are attached with stycast epoxy to a thermally isolating polymer structure made of SU-8. SU-8 is a negative epoxy based photo resist produced by Microchem. We describe the fabrication of the X-ray absorbers and their suspension on a handle die in an adhesive matrix. We describe the production process for the polymer isolators on the detector elements. We have developed a new process for the alignment, and simultaneous bonding of the absorber tiles to an entire detector array. This process uses equipment and techniques used in the flip-chip bonding industry and approaches developed in the fabrication of the XRS-2 instrument. XRS-2 was an X-ray spectrometer that was launched on the Suzaku telescope in July 10, 2005. We describe the process and show examples of sample arrays produced by this process. Arrays with up to 300 elements have been bonded. The present tests have used dummy absorbers made of Si. In future work, we will demonstrate bonding of HgTe absorbers.

Moseley, S.↗

The Design, Implementation, and Performance of the Astro-H SXS Aperture Assembly and Blocking Filters

The calorimeter array of the JAXA Astro-H (renamed Hitomi) Soft X-ray Spectrometer (SXS) was designed to provide unprecedented spectral resolution of spatially extended cosmic x-ray sources and of all cosmic x-ray sources in the Fe-K band around 6 keV. The properties that make the SXS a powerful x-ray spectrometer also make it sensitive to the entire electromagnetic band. If characterized as a bolometer, it would have a noise equivalent power (NEP) of < 4x10(exp -18) W/(Hz)0.5. Thus it was imperative to shield the detector from thermal radiation from the instrument and optical and UV photons from the sky. Additionally, it was necessary to shield the coldest stages of the instrument from the thermal radiation emanating from the warmer stages. These needs are addressed by a series of five thin-film radiation blocking filters that block long-wavelength radiation while minimizing x-ray attenuation. The SXS aperture assembly is a system of barriers, baffles, filter carriers, and filter mounts that supports the filters and inhibits their potential contamination. The three warmer filters also were equipped with thermometers and heaters for decontamination.

microcalorimeter↗

Vibration Isolation System for Cryocoolers of Soft X-Ray Spectrometer (SXS) Onboard ASTRO-H (Hitomi)

Soft X-ray Spectrometer (SXS) onboard ASTRO-H (named Hitomi after launch) is a micro-calorimeter-type spectrometer, installed in a dewar to be cooled at 50 mK. The energy resolution of the SXS engineering model suffered from micro-vibration from cryocoolers mounted on the dewar. This is mitigated for the flight model by introducing vibration isolation systems between the cryocoolers and the dewar. The detector performance of the flight model was verified before launch of the spacecraft in both ambient condition and thermal-vac condition, showing no detectable degradation in energy resolution. The in-orbit performance was also consistent with that on ground, indicating that the cryocoolers were not damaged by launch environment. The design and performance of the vibration isolation system along with the mechanism of how the micro-vibration could degrade the cryogenic detector is shown.

Microcalorimeter↗