The V-3 contamination test of the chamber A facility and a subsequent cryogenic/vacuum study of the V-3 test quartz crystal microbalance
The areas of orbital and ground contamination of flight experiment hardware have been well established. This report relates directly to results of vacuum chamber testing for the ground evaluation of flight experiment hardware performance. First, the data obtained during the V-3 contamination testing in the Johnson Space Center's Chamber A space simulation test facility are presented. Second, during the V-3 contamination tests, the MSFC Space Sciences Laboratory's quartz crystal microbalance exhibited two periods of anomalous readings. Therefore, a subsequent small chamber tests was conducted in a controlled cryogenic/vacuum environment. The objective was to reproduce with known parameters the anomalous behavior patterns of the V-3 test data. Analyses of the anomalous readings are made on the basis of these tests. Additionally, as a by-product of the small chamber tests, calibration curves then existing for the quartz crystal microbalance were empirically extended, and certain data-formatting aids were documented.