Resonance Counters as the Best Tool for the Investigations in Material Science
Sensitivity and resolution play a crucial role when Moessbauer spectroscopy is used in the materials science. Application of resonance counters in Moessbauer spectrometers allows us to increase the parameters mentioned above, and also signal-to-noise ratio considerably. The last one provides diminishing the time needed for obtaining given statistical accuracy. We carried out investigations of development of optimal counters for following isotopes: {sup 57}Fe, {sup 119}Sn, and {sup 151}Eu. Influence of different parameters of resonant radiation converters on experimental results was considered theoretically. Optimization of design has been performed using mathematical modeling based on Monte-Carlo method. Comparison of different types of counters used for resonant detecting was carried out. Results of experimental works on selection of efficient radiation converters are given. Comparison of scintillation and gas resonance counters was carried out. FeAl and FeGe{sub 2} alloys and K{sub 2}MgFe(CN){sub 6} have been used as converters for experiments with {sup 57}Fe-isotope, CaSnO{sub 3} has been used for {sup 119}Sn and Eu{sub 2}O{sub 3} and EuF{sub 3}--for {sup 151}Eu isotope. Gamma-optical scheme for versatile spectrometer, which expands the range of application of resonant detection for other Moessbauer isotopes, was suggested.