Poster — Thur Eve — 22: A water calorimeter for low-energy particle beams
In this work, the feasibility of absolute dose to water measurements in low-energy electron beams using a water calorimeter specifically developed for shallow measurements is established. The calorimeter design consists of a cylindrical glass vessel encased in a block of expanded polystyrene. The vessel has a front window thickness of 1.1 mm, a 4 cm radius, and is 2.5 cm in depth. The vessel-block assembly sits inside a thermally-insulated box and is air-cooled to an operating temperature of 4 °C. Radiation-induced thermal gradients were simulated in a geometric model of the calorimeter using a finite element analysis software package. 52 absorbed dose to water measurements were performed in a 6 and 8 MeV electron beam (z{sub max} of 1.32 and 1.76 cm, respectively) for 60 seconds at a repetition rate of 400 MU/min and an SSD of 120 cm. Within the vessel, the depth of measurement was set to 1.08 cm relative to the inner front window. The average measured dose to water was 59.6 ± 0.2 cGy/100 MU (6 MeV), and 63.7 ± 0.3 cGy/100 MU (8 MeV). The associated heat transfer corrections were determined to be 1.026 ± 0.003 and 1.017 ± 0.004 for the 6 and 8 MeV beams, respectively. The most significant source of uncertainty in this study was the repeatability (type A, 0.42%). It is expected that performing fewer consecutive measurements under higher dose rate conditions will improve the stability of the thermal background, thereby improving the repeatability and reducing the overall standard uncertainty.