Evaluation of Sampling Inlets for Continuous Air Monitors in Calm Air Conditions
The Canberra NetCAM (networked continuous air monitor) can be operated with a radial inlet attachment from the manufacturer. Sampled aerosol can be transported through large diameter tubing from a desired location to the NetCAM sampler. In this assessment, the effect of the addition of the radial inlet was investigated. Also, a sampling system consisting of a vertical sample tube and a 90-degree elbow (to the horizontal plane) was evaluated by empirical calculations with the DEPO software. Generalized aspiration ratio calculations were done with the empirical model from Vincent. Three different sample probe types were also experimentally tested. In summary: (1) For typical room air conditions, the addition of the Canberra inline adapter to the NetCAM has no discernible effect on system performance. This was determined by literature search with a previously published study. (2) The DEPO software (US NRC / Texas A&M University) was used to show the effect of adding a 2-meter high vertical tube and a ninety-degree tube bend to a sampling inlet. In this example, there was a total penetration of 97.7% of 10-micron diameter aerosol particles for an air flowrate of 2.0 ACFM (actual cubic feet per minute). (3) Calculations of nozzle aspiration ratio (for general, non-specific situations) were performed using the Vincent empirical model. For example, in typical room air, where the average velocity is less than 0.5 m/sec, an aerosol inlet probe has a calculated aspiration ratio (or penetration efficiency) value between 98 and 100%, depending on the angle of the probe inlet compared to the direction of the incoming air velocity vector. A table of nozzle aspiration ratios was generated for a range of different room air velocities and probe inlet angles. (4) Three different candidate probe geometries were tested, (A) a probe with a vertical entry and vertical transport tube, (B) a probe with a horizontal entry and vertical transport tube, and (C) a ‘widemouth’ probe with a vertical entry and vertical transport tube. The experimentally determined values were 98%, 102%, and 100%, respectively, for the aerosol penetration into these inlets. In conclusion, the addition of a complete sampling system with probe inlet, a short length of sampling lines and the radial inlet would still be able to deliver about 95% of the sampled aerosol from the sampling location to the Canberra NetCAM.