Multiple Coulomb scattering in acrylic of a 221.3 MeV therapeutic proton beam
Measurements of multiple Coulomb scattering (MCS) distributions for 221.3 MeV therapeutic protons are presented using a novel detector system comprised of a thin scintillator, a pellicle mirror, and a digital camera. The MCS distributions were characterized for three acrylic phantoms of varying lengths and for two biological density-equivalent phantoms simulating bone and muscle. Additionally, beam profiles were measured across an energy range of 80.3–221.3 MeV in 20 MeV increments. The observed energy dependence of the photon yields is consistent with the tabulated stopping power values. Finally, the experimental results are benchmarked against Geant4 simulations, demonstrating consistent agreement and validating the capability of the detector system for radiology measurements.