At least 127 records · Page 7

WE-G-BRC-02: Risk Assessment for HDR Brachytherapy

Failure Mode and Effects Analysis (FMEA) originated as an industrial engineering technique used for risk management and safety improvement of complex processes. In the context of radiotherapy, the AAPM Task Group 100 advocates FMEA as the framework of choice for establishing clinical quality management protocols. However, there is concern that widespread adoption of FMEA in radiation oncology will be hampered by the perception that implementation of the tool will have a steep learning curve, be extremely time consuming and labor intensive, and require additional resources. To overcome these preconceptions and facilitate the introduction of the tool into clinical practice, the medical physics community must be educated in the use of this tool and the ease in which it can be implemented. Organizations with experience in FMEA should share their knowledge with others in order to increase the implementation, effectiveness and productivity of the tool. This session will include a brief, general introduction to FMEA followed by a focus on practical aspects of implementing FMEA for specific clinical procedures including HDR brachytherapy, physics plan review and radiosurgery. A description of common equipment and devices used in these procedures and how to characterize new devices for safe use in patient treatments will be presented. This will be followed by a discussion of how to customize FMEA techniques and templates to one’s own clinic. Finally, cases of common failure modes for specific procedures (described previously) will be shown and recommended intervention methodologies and outcomes reviewed. Learning Objectives: Understand the general concept of failure mode and effect analysis Learn how to characterize new equipment for safety Be able to identify potential failure modes for specific procedures and learn mitigation techniques Be able to customize FMEA examples and templates for use in any clinic.

60 APPLIED LIFE SCIENCES↗

WE-G-BRC-03: Risk Assessment for Physics Plan Review

Failure Mode and Effects Analysis (FMEA) originated as an industrial engineering technique used for risk management and safety improvement of complex processes. In the context of radiotherapy, the AAPM Task Group 100 advocates FMEA as the framework of choice for establishing clinical quality management protocols. However, there is concern that widespread adoption of FMEA in radiation oncology will be hampered by the perception that implementation of the tool will have a steep learning curve, be extremely time consuming and labor intensive, and require additional resources. To overcome these preconceptions and facilitate the introduction of the tool into clinical practice, the medical physics community must be educated in the use of this tool and the ease in which it can be implemented. Organizations with experience in FMEA should share their knowledge with others in order to increase the implementation, effectiveness and productivity of the tool. This session will include a brief, general introduction to FMEA followed by a focus on practical aspects of implementing FMEA for specific clinical procedures including HDR brachytherapy, physics plan review and radiosurgery. A description of common equipment and devices used in these procedures and how to characterize new devices for safe use in patient treatments will be presented. This will be followed by a discussion of how to customize FMEA techniques and templates to one’s own clinic. Finally, cases of common failure modes for specific procedures (described previously) will be shown and recommended intervention methodologies and outcomes reviewed. Learning Objectives: Understand the general concept of failure mode and effect analysis Learn how to characterize new equipment for safety Be able to identify potential failure modes for specific procedures and learn mitigation techniques Be able to customize FMEA examples and templates for use in any clinic.

60 APPLIED LIFE SCIENCES↗

Independent Orbiter Assessment (IOA): Analysis of the guidance, navigation, and control subsystem

The results of the Independent Orbiter Assessment (IOA) of the Failure Modes and Effects Analysis (FMEA) and Critical Items List (CIL) is presented. The IOA approach features a top-down analysis of the hardware to determine failure modes, criticality, and potential critical items. To preserve independence, this analysis was accomplished without reliance upon the results contained within the NASA FMEA/CIL documentation. The independent analysis results corresponding to the Orbiter Guidance, Navigation, and Control (GNC) Subsystem hardware are documented. The function of the GNC hardware is to respond to guidance, navigation, and control software commands to effect vehicle control and to provide sensor and controller data to GNC software. Some of the GNC hardware for which failure modes analysis was performed includes: hand controllers; Rudder Pedal Transducer Assembly (RPTA); Speed Brake Thrust Controller (SBTC); Inertial Measurement Unit (IMU); Star Tracker (ST); Crew Optical Alignment Site (COAS); Air Data Transducer Assembly (ADTA); Rate Gyro Assemblies; Accelerometer Assembly (AA); Aerosurface Servo Amplifier (ASA); and Ascent Thrust Vector Control (ATVC). The IOA analysis process utilized available GNC hardware drawings, workbooks, specifications, schematics, and systems briefs for defining hardware assemblies, components, and circuits. Each hardware item was evaluated and analyzed for possible failure modes and effects. Criticality was assigned based upon the severity of the effect for each failure mode.

Trahan, W. H.↗

Probabilistic Risk Assessment Procedures Guide for NASA Managers and Practitioners (Second Edition)

Probabilistic Risk Assessment (PRA) is a comprehensive, structured, and logical analysis method aimed at identifying and assessing risks in complex technological systems for the purpose of cost-effectively improving their safety and performance. NASA's objective is to better understand and effectively manage risk, and thus more effectively ensure mission and programmatic success, and to achieve and maintain high safety standards at NASA. NASA intends to use risk assessment in its programs and projects to support optimal management decision making for the improvement of safety and program performance. In addition to using quantitative/probabilistic risk assessment to improve safety and enhance the safety decision process, NASA has incorporated quantitative risk assessment into its system safety assessment process, which until now has relied primarily on a qualitative representation of risk. Also, NASA has recently adopted the Risk-Informed Decision Making (RIDM) process [1-1] as a valuable addition to supplement existing deterministic and experience-based engineering methods and tools. Over the years, NASA has been a leader in most of the technologies it has employed in its programs. One would think that PRA should be no exception. In fact, it would be natural for NASA to be a leader in PRA because, as a technology pioneer, NASA uses risk assessment and management implicitly or explicitly on a daily basis. NASA has probabilistic safety requirements (thresholds and goals) for crew transportation system missions to the International Space Station (ISS) [1-2]. NASA intends to have probabilistic requirements for any new human spaceflight transportation system acquisition. Methods to perform risk and reliability assessment in the early 1960s originated in U.S. aerospace and missile programs. Fault tree analysis (FTA) is an example. It would have been a reasonable extrapolation to expect that NASA would also become the world leader in the application of PRA. That was, however, not to happen. Early in the Apollo program, estimates of the probability for a successful roundtrip human mission to the moon yielded disappointingly low (and suspect) values and NASA became discouraged from further performing quantitative risk analyses until some two decades later when the methods were more refined, rigorous, and repeatable. Instead, NASA decided to rely primarily on the Hazard Analysis (HA) and Failure Modes and Effects Analysis (FMEA) methods for system safety assessment.

Stamatelatos,Michael↗

Probabilistic Risk Assessment: A Bibliography

Probabilistic risk analysis is an integration of failure modes and effects analysis (FMEA), fault tree analysis and other techniques to assess the potential for failure and to find ways to reduce risk. This bibliography references 160 documents in the NASA STI Database that contain the major concepts, probabilistic risk assessment, risk and probability theory, in the basic index or major subject terms, An abstract is included with most citations, followed by the applicable subject terms.

Source record↗

Reliability analysis based on operational success criteria

Failure Modes, Effects, and Criticality Analysis Model disregards failures of insignificant components. Assumptions used to establish baselines for reliability prediction and analytical method are listed. Techniques are applicable to any commercial design and can be extended into software or management systems.

Esmond, F. G.↗

SU-F-T-243: Major Risks in Radiotherapy. A Review Based On Risk Analysis Literature

Purpose: We present a literature review of risk analyses in radiotherapy to highlight the most reported risks and facilitate the spread of this valuable information so that professionals can be aware of these major threats before performing their own studies. Methods: We considered studies with at least an estimation of the probability of occurrence of an adverse event (O) and its associated severity (S). They cover external beam radiotherapy, brachytherapy, intraoperative radiotherapy, and stereotactic techniques. We selected only the works containing a detailed ranked series of elements or failure modes and focused on the first fully reported quartile as much. Afterward, we sorted the risk elements according to a regular radiotherapy procedure so that the resulting groups were cited in several works and be ranked in this way. Results: 29 references published between 2007 and February 2016 were studied. Publication trend has been generally rising. The most employed analysis has been the Failure mode and effect analysis (FMEA). Among references, we selected 20 works listing 258 ranked risk elements. They were sorted into 31 groups appearing at least in two different works. 11 groups appeared in at least 5 references and 5 groups did it in 7 or more papers. These last sets of risks where choosing another set of images or plan for planning or treating, errors related with contours, errors in patient positioning for treatment, human mistakes when programming treatments, and planning errors. Conclusion: There is a sufficient amount and variety of references for identifying which failure modes or elements should be addressed in a radiotherapy department before attempting a specific analysis. FMEA prevailed, but other studies such as “risk matrix” or “occurrence × severity” analyses can also lead professionals’ efforts. Risk associated with human actions ranks very high; therefore, they should be automated or at least peer-reviewed.

60 APPLIED LIFE SCIENCES↗

CalWave Open Water Demo - FMEA Update Budget Period 2

The Failure Modes and Effects Analysis (FMEA) is a qualitative reliability technique for systematically analyzing each possible failure mode within a hardware system, and identifying the resulting effect on that system, the mission, and the personnel. This submission includes an updated FMEA summary for CalWave's open water demonstration including pre- and post-mitigation results, hazard identification (HAZID) analysis, and component/function rooted FMEA.

16 TIDAL AND WAVE POWER↗

Spaceflight Ground Support Equipment Reliability & System Safety Data

Presented were Reliability Analysis, consisting primarily of Failure Modes and Effects Analysis (FMEA), and System Safety Analysis, consisting of Preliminary Hazards Analysis (PHA), performed to ensure that the CoNNeCT (Communications, Navigation, and Networking re- Configurable Testbed) Flight System was safely and reliably operated during its Assembly, Integration and Test (AI&T) phase. A tailored approach to the NASA Ground Support Equipment (GSE) standard, NASA-STD-5005C, involving the application of the appropriate Requirements, S&MA discipline expertise, and a Configuration Management system (to retain a record of the analysis and documentation) were presented. Presented were System Block Diagrams of selected GSE and the corresponding FMEA, as well as the PHAs. Also discussed are the specific examples of the FMEAs and PHAs being used during the AI&T phase to drive modifications to the GSE (via "redlining" of test procedures, and the placement of warning stickers to protect the flight hardware) before being interfaced to the Flight System. These modifications were necessary because failure modes and hazards were identified during the analysis that had not been properly mitigated. Strict Configuration Management was applied to changes (whether due to upgrades or expired calibrations) in the GSE by revisiting the FMEAs and PHAs to reflect the latest System Block Diagrams and Bill Of Material. The CoNNeCT flight system has been successfully assembled, integrated, tested, and shipped to the launch site without incident. This demonstrates that the steps taken to safeguard the flight system when it was interfaced to the various GSE were successful.

Fernandez, Rene↗

Independent Orbiter Assessment (IOA): Assessment of the mechanical actuation subsystem, volume 2

The results of the Independent Orbiter Assessment (IOA) of the Failure Modes and Effects Analysis (FMEA) and Critical Items List (CIL) are presented. The IOA approach features a top-down analysis of the hardware to determine draft failure modes, criticality, and potential critical items. To preserve independence, this analysis was accomplished without reliance upon the results contained within the NASA FMEA/CIL documentation. The IOA results were then compared to the proposed Post 51-L NASA FMEA/CIL baseline that was available. A resolution of each discrepancy from the comparison was provided through additional analysis as required. These discrepancies were flagged as issues, and recommendations were made based on the FMEA data available at the time. This report documents the results of that comparison for the Orbiter Mechanical Actuation System (MAS) hardware. Specifically, the MAS hardware consists of the following components: Air Data Probe (ADP); Elevon Seal Panel (ESP); External Tank Umbilical (ETU); Ku-Band Deploy (KBD); Payload Bay Doors (PBD); Payload Bay Radiators (PBR); Personnel Hatches (PH); Vent Door Mechanism (VDM); and Startracker Door Mechanism (SDM). Criticality was assigned based upon the severity of the effect for each failure mode. Volume 2 continues the presentation of IOA analysis worksheets and contains the potential critical items list, detailed analysis, and NASA FMEA/CIL to IOA worksheet cross reference and recommendations.

Bradway, M. W.↗

Generation and analysis of large reliability models

An effort has been underway for several years at NASA's Langley Research Center to extend the capability of Markov modeling techniques for reliability analysis to the designers of highly reliable avionic systems. This effort has been focused in the areas of increased model abstraction and increased computational capability. The reliability model generator (RMG), a software tool which uses as input a graphical, object-oriented block diagram of the system, is discussed. RMG uses an automated failure modes-effects analysis algorithm to produce the reliability model from the graphical description. Also considered is the ASSURE software tool, a parallel processing program which uses the ASSIST modeling language and SURE semi-Markov solution technique. An executable failure modes-effects analysis is used by ASSURE. The successful combination of the power of graphical representation, automated model generation, and parallel computation leads to the conclusion that large system architectures can now be analyzed.

Aerospace electronics↗

Quality Assurance With Plan Veto: Reincarnation of a Record and Verify System and Its Potential Value

Purpose: To quantify the potential impact of the Integrating the Healthcare Enterprise–Radiation Oncology Quality Assurance with Plan Veto (QAPV) on patient safety of external beam radiation therapy (RT) operations. Methods and Materials: An institutional database of events (errors and near-misses) was used to evaluate the ability of QAPV to prevent clinically observed events. We analyzed reported events that were related to Digital Imaging and Communications in Medicine RT plan parameter inconsistencies between the intended treatment (on the treatment planning system) and the delivered treatment (on the treatment machine). Critical Digital Imaging and Communications in Medicine RT plan parameters were identified. Each event was scored for importance using the Failure Mode and Effects Analysis methodology. Potential error occurrence (frequency) was derived according to the collected event data, along with the potential event severity, and the probability of detection with and without the theoretical implementation of the QAPV plan comparison check. Failure Mode and Effects Analysis Risk Priority Numbers (RPNs) with and without QAPV were compared to quantify the potential benefit of clinical implementation of QAPV. Results: The implementation of QAPV could reduce the RPN values for 15 of 22 (71%) of evaluated parameters, with an overall average reduction in RPN of 68 (range, 0-216). For the 6 high-risk parameters (>200), the average reduction in RPN value was 163 (range, 108-216). The RPN value reduction for the intermediate-risk (200 > RPN > 100) parameters was (0-140). With QAPV, the largest RPN value for “Beam Meterset” was reduced from 324 to 108. The maximum reduction in RPN value was for Beam Meterset (216, 66.7%), whereas the maximum percentage reduction was for Cumulative Meterset Weight (80, 88.9%). Conclusion: This analysis quantifies the value of the Integrating the Healthcare Enterprise–Radiation Oncology QAPV implementation in clinical workflow. We demonstrate that although QAPV does not provide a comprehensive solution for error prevention in RT, it can have a significant impact on a subset of the most severe clinically observed events.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Independent Orbiter Assessment (IOA): Analysis of the instrumentation subsystem

The results of the Independent Orbiter Assessment (IOA) of the Failure Modes and Effects Analysis (FMEA) and Critical Items List (CIL) are presented. The IOA approach features a top-down analysis of the hardware to determine failure modes, criticality, and potential critical items. To preserve independence, this analysis was accomplished without reliance upon the results contained within the NASA FMEA/CIL documentation. The independent analysis results for the Instrumentation Subsystem are documented. The Instrumentation Subsystem (SS) consists of transducers, signal conditioning equipment, pulse code modulation (PCM) encoding equipment, tape recorders, frequency division multiplexers, and timing equipment. For this analysis, the SS is broken into two major groupings: Operational Instrumentation (OI) equipment and Modular Auxiliary Data System (MADS) equipment. The OI equipment is required to acquire, condition, scale, digitize, interleave/multiplex, format, and distribute operational Orbiter and payload data and voice for display, recording, telemetry, and checkout. It also must provide accurate timing for time critical functions for crew and payload specialist use. The MADS provides additional instrumentation to measure and record selected pressure, temperature, strain, vibration, and event data for post-flight playback and analysis. MADS data is used to assess vehicle responses to the flight environment and to permit correlation of such data from flight to flight. The IOA analysis utilized available SS hardware drawings and schematics for identifying hardware assemblies and components and their interfaces. Criticality for each item was assigned on the basis of the worst-case effect of the failure modes identified.

Howard, B. S.↗

Reliability of EUCLIDIAN: An autonomous robotic system for image-guided prostate brachytherapy

Purpose: Recently, several robotic systems have been developed to perform accurate and consistent image-guided brachytherapy. Before introducing a new device into clinical operations, it is important to assess the reliability and mean time before failure (MTBF) of the system. In this article, the authors present the preclinical evaluation and analysis of the reliability and MTBF of an autonomous robotic system, which is developed for prostate seed implantation. Methods: The authors have considered three steps that are important in reliability growth analysis. These steps are: Identification and isolation of failures, classification of failures, and trend analysis. For any one-of-a-kind product, the reliability enhancement is accomplished through test-fix-test. The authors have used failure mode and effect analysis for collection and analysis of reliability data by identifying and categorizing the failure modes. Failures were classified according to severity. Failures that occurred during the operation of this robotic system were considered as nonhomogenous Poisson process. The failure occurrence trend was analyzed using Laplace test. For analyzing and predicting reliability growth, commonly used and widely accepted models, Duane's model and the Army Material Systems Analysis Activity, i.e., Crow's model, were applied. The MTBF was used as an important measure for assessing the system's reliability. Results: During preclinical testing, 3196 seeds (in 53 test cases) were deposited autonomously by the robot and 14 critical failures were encountered. The majority of the failures occurred during the first few cases. The distribution of failures followed Duane's postulation as well as Crow's postulation of reliability growth. The Laplace test index was -3.82 (<0), indicating a significant trend in failure data, and the failure intervals lengthened gradually. The continuous increase in the failure occurrence interval suggested a trend toward improved reliability. The MTBF was 592 seeds, which implied that several prostate seed implantation cases would be possible without encountering any critical failure. The shape parameter for the MTBF was 0.3859 (<1), suggesting a positive reliability growth of this robotic system. At 95% confidence, the reliability for deposition of 65 seeds was more than 90%. Conclusions: Analyses of failure mode strongly indicated a gradual improvement of reliability of this autonomous robotic system. High MTBF implied that several prostate seed implant cases would be possible without encountering any critical failure.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

ALS rocket engine combustion devices design and demonstration

Work performed during Phase one is summarized and the significant technical and programmatic accomplishments occurring during this period are documented. Besides a summary of the results, methodologies, trade studies, design, fabrication, and hardware conditions; the following are included: the evolving Maintainability Plan, Reliability Program Plan, Failure Summary and Analysis Report, and the Failure Mode and Effect Analysis.

Arreguin, Steve↗

Systems Health Monitoring: Integrating FMEA into Bayesian Networks

The foreseeable high traffic density suggests that a large number of electric propulsion systems will enter the airspace, and that they will also operate at high frequency, e.g., large number of take offs and landings per unit time. The reliability of such critical systems is therefore key to ensure high safety standards in the low-altitude airspace. Diagnostic systems, which aim at identifying incipient faults, can mitigate unexpected failures or lower-than-expected reliability by performing early fault detection by monitoring the systems. A key element of fault diagnosis is fault detection and isolation (FDI), which complexity increases with the complexity of the system itself, namely the number of subsystems and components, interactions among sub-systems, and the number of sensors available. The proposed approach leverages combination of failure mode and effect analysis (FMEA) integrated with Bayesian networks, thus introducing dependability structures into a diagnostic framework to aid FDI. Faults and failure events from the FMEA are mapped within a Bayesian network, where network edges replicate the links embedded within FMEAs. The integrated framework enables the fault isolation process by identifying the probability of occurrence of specific faults or root causes given evidence observed through sensor signals. In this work, sub systems of Urban Air Mobility (UAM) type vehicle like avionics, structures, power-train etc. are taken into account to show the approach at the system level. This work integrates early design phase in the development of UAM type vehicles with diagnostic tools, which are often developed later in the product life-cycle, or retrofitted at a later time on systems. Failure mode and effect analysis (FMEA) derived for the system in the design phase is embedded within a Bayesian network (BN).

UAM↗

A computer aided treatment event recognition system in radiation therapy

Purpose: To develop an automated system to safeguard radiation therapy treatments by analyzing electronic treatment records and reporting treatment events. Methods: CATERS (Computer Aided Treatment Event Recognition System) was developed to detect treatment events by retrieving and analyzing electronic treatment records. CATERS is designed to make the treatment monitoring process more efficient by automating the search of the electronic record for possible deviations from physician's intention, such as logical inconsistencies as well as aberrant treatment parameters (e.g., beam energy, dose, table position, prescription change, treatment overrides, etc). Over a 5 month period (July 2012–November 2012), physicists were assisted by the CATERS software in conducting normal weekly chart checks with the aims of (a) determining the relative frequency of particular events in the authors’ clinic and (b) incorporating these checks into the CATERS. During this study period, 491 patients were treated at the University of Iowa Hospitals and Clinics for a total of 7692 fractions. Results: All treatment records from the 5 month analysis period were evaluated using all the checks incorporated into CATERS after the training period. About 553 events were detected as being exceptions, although none of them had significant dosimetric impact on patient treatments. These events included every known event type that was discovered during the trial period. A frequency analysis of the events showed that the top three types of detected events were couch position override (3.2%), extra cone beam imaging (1.85%), and significant couch position deviation (1.31%). The significant couch deviation is defined as the number of treatments where couch vertical exceeded two times standard deviation of all couch verticals, or couch lateral/longitudinal exceeded three times standard deviation of all couch laterals and longitudinals. On average, the application takes about 1 s per patient when executed on either a desktop computer or a mobile device. Conclusions: CATERS offers an effective tool to detect and report treatment events. Automation and rapid processing enables electronic record interrogation daily, alerting the medical physicist of deviations potentially days prior to performing weekly check. The output of CATERS could also be utilized as an important input to failure mode and effects analysis.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Independent Orbiter Assessment (IOA): Analysis of the Orbiter Experiment (OEX) subsystem

The results of the Independent Orbiter Assessment (IOA) of the Failure Modes and Effects Analysis (FMEA) and Critical Items List (CIL) are presented. The IOA approach features a top-down analysis of the hardware to determine failure modes, criticality, and potential critical items. To preserve independence, this analysis was accomplished without reliance upon the results contained within the NASA FMEA/CIL documentation. This report documents the independent analysis results corresponding to the Orbiter Experiments hardware. Each level of hardware was evaluated and analyzed for possible failure modes and effects. Criticality was assigned based upon the severity of the effect for each failure mode. The Orbiter Experiments (OEX) Program consists of a multiple set of experiments for the purpose of gathering environmental and aerodynamic data to develop more accurate ground models for Shuttle performance and to facilitate the design of future spacecraft. This assessment only addresses currently manifested experiments and their support systems. Specifically this list consists of: Shuttle Entry Air Data System (SEADS); Shuttle Upper Atmosphere Mass Spectrometer (SUMS); Forward Fuselage Support System for OEX (FFSSO); Shuttle Infrared Laced Temperature Sensor (SILTS); Aerodynamic Coefficient Identification Package (ACIP); and Support System for OEX (SSO). There are only two potential critical items for the OEX, since the experiments only gather data for analysis post mission and are totally independent systems except for power. Failure of any experiment component usually only causes a loss of experiment data and in no way jeopardizes the crew or mission.