Relaxation measurements of an MRI system phantom at low magnetic field strengths
Abstract Objective Temperature controlled T 1 and T 2 relaxation times are measured on NiCl 2 and MnCl 2 solutions from the ISMRM/NIST system phantom at low magnetic field strengths of 6.5 mT, 64 mT and 550 mT. Materials and methods The T 1 and T 2 were measured of five samples with increasing concentrations of NiCl 2 and five samples with increasing concentrations of MnCl 2 . All samples were scanned at 6.5 mT, 64 mT and 550 mT, at sample temperatures ranging from 10 °C to 37 °C. Results The NiCl 2 solutions showed little change in T 1 and T 2 with magnetic field strength, and both relaxation times decreased with increasing temperature. The MnCl 2 solutions showed an increase in T 1 and a decrease in T 2 with increasing magnetic field strength, and both T 1 and T 2 increased with increasing temperature. Discussion The low field relaxation rates of the NiCl 2 and MnCl 2 arrays in the ISMRM/NIST system phantom are investigated and compared to results from clinical field strengths of 1.5 T and 3.0 T. The measurements can be used as a benchmark for MRI system functionality and stability, especially when MRI systems are taken out of the radiology suite or laboratory and into less traditional environments.