An improved offset generator developed for Allan deviation measurement of ultra stable frequency standards
Measuring the performance of ultra stable frequency standards such as the Superconducting Cavity Maser Oscillator (SCMO) necessitates improvement of some test instrumentation. The frequency stability test equipment used at JPL includes a 1 Hz Offset Generator to generate a beat frequency between a pair of 100 MHz signals that are being compared. The noise floor of the measurement system using the current Offset Generator is adequate to characterize stability of hydrogen masers, but it is not adequate for the SCMO. A new Offset Generator with improved stability was designed and tested at JPL. With this Offset Generator and a new Zero Crossing Detector, recently developed at JPL, the measurement flow was reduced by a factor of 5.5 at 1 second tau, 3.0 at 1000 seconds, and 9.4 at 10,000 seconds, compared against the previous design. In addition to the new circuit designs of the Offset Generator and Zero Crossing Detector, tighter control of the measurement equipment environment was required to achieve this improvement. The design of this new Offset Generator are described, along with details of the environment control methods used.