Induced environment contamination monitor (IECM) cascade impactor
The Cascade Impactor on the IECM is an aerosol sampling instrument which employs three cascade impaction-type nozzles of progressively smaller diameter. The considered approach makes it possible to achieve particle size discrimination involving three ranges (5 microns and larger, 1 to 5 microns, and 0.3 to 1 micron). In addition to the particulate measuring cascade stages, a fourth quartz crystal microbalance measures the cumulative airborne molecular nonvolatile residue (NVR) at ambient temperature. The cascade measurements provide a temporal record of the payload bay's particulate environment during ascent, descent, and immediate postlanding periods. The NVR stage functions throughout the mission. Particulates of a size greater than five microns in diameter were below 10 micrograms per cu m throughout the mission.