Surface Metrology for X-ray Mirrors using Deflectometry (Final Report)
Explore the source record for details and available documents.
SEARCH · Search NASA
Search indexed NASA NTRS and DOE OSTI research on propulsion, heat transfer, battery materials and energy systems. Follow report and document links to the original sources.
Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.
Explore the source record for details and available documents.
The Software Configurable Optical Test System (SCOTS) uses deflectometry to measure surface slopes of general optical shapes without the need for additional null optics. Careful alignment of test geometry and calibration of inherent system error improve the accuracy of SCOTS to a level where it competes with interferometry. We report a SCOTS surface measurement of an off-axis superpolished elliptical x-ray mirror that achieves <1 nm<1 nm root-mean-square accuracy for the surface measurement with low-order term included.
Slope measuring deflectometry is commonly used by the X-ray optics community to measure the long-spatial-wavelength surface figure error of optical components dedicated to guide and focus X-rays under grazing incidence condition at synchrotron and free electron laser beamlines. The best performing instruments of this kind are capable of absolute accuracy on the level of 30-50 nrad. However, the exact bandwidth of the measurements, determined at the higher spatial frequencies by the instrument’s spatial resolution, or more generally by the instrument’s modulation transfer function (MTF) is hard to determine. An MTF calibration method based on application of a test surface with a one-dimensional (1D) chirped height profile of constant amplitude was suggested in the past. In this work, we propose a new approach to designing the test surfaces with a 2D-chirped topography, specially optimized for MTF characterization of slope measuring instruments. The design of the developed MTF test samples based on the proposed linear chirped slope profiles (LCSPs) is free of the major drawback of the 1D chirped height profiles, where in the slope domain, the amplitude strongly increases with the local spatial frequency of the profile. We provide the details of fabrication of the LCSP samples. The results of first application of the developed test samples to measure the spatial resolution of the BESSY-NOM at different experimental arrangements are also presented and discussed.
Systematic error and instrumental drift are the major limiting factors of sub-microradian slope metrology with state-of-the-art x-ray optics. Significant suppression of the errors can be achieved by using an optimal measurement strategy suggested in [Rev. Sci. Instrum. 80, 115101 (2009)]. With this series of LSBL Notes, we report on development of an automated, kinematic, rotational system that provides fully controlled flipping, tilting, and shifting of a surface under test. The system is integrated into the Advanced Light Source long trace profiler, LTP-II, allowing for complete realization of the advantages of the optimal measurement strategy method. We provide details of the system?s design, operational control and data acquisition. The high performance of the system is demonstrated via the results of high precision measurements with a spherical test mirror.
Systematic error and instrumental drift are the major limiting factors of sub-microradian slope metrology with state-of-the-art x-ray optics. Significant suppression of the errors can be achieved by using an optimal measurement strategy suggested in [Rev. Sci. Instrum. 80, 115101 (2009)]. With this series of LSBL Notes, we report on development of an automated, kinematic, rotational system that provides fully controlled flipping, tilting, and shifting of a surface under test. The system is integrated into the Advanced Light Source long trace profiler, LTP-II, allowing for complete realization of the advantages of the optimal measurement strategy method. We provide details of the system’s design, operational control and data acquisition. The high performance of the system is demonstrated via the results of high precision measurements with a spherical test mirror.
Here, we investigate and compare the spatial (lateral) resolution, or more generally, the optical/instrumental transfer function (OTF/ITF) of surface slope measuring profilometers of two different types that are commonly used for high accuracy characterization of x-ray optics at the long-spatial-wavelength range. These are an autocollimator based profiler, Optical Surface Measuring System (OSMS), and a long trace profiler, LTP-II, both available at the Advanced Light Source (ALS) X-Ray Optics Lab (XROL). In the OSMS, an ELCOMAT-3000 electronic auto-collimator, vertically mounted to the translation carriage and equipped with an aperture of 2.5 mm diameter, is scanned along the surface under test. The LTP-II OTF has been measured for two different configurations, a classical two-beam pencil-beam-interferometry and a single-Gaussian-beam deflectometry. For the ITF calibration, we apply a recently developed method based on test surfaces with one-dimensional (1D) linear chirped height profiles of constant slope amplitude. Analytical expressions for the OTFs, empirically deduced based on the experimental results, are presented. We also discuss the application of the results of the ITF measurements and modeling to improve the surface slope metrology with state-of-the-art x-ray mirrors.
The proliferation of autocollimator-based surface profilometers at synchrotron metrology laboratories worldwide necessitates a detailed understanding of the parameters influencing their angular response. A comprehensive overview of the current status of autocollimator characterization and calibration at the Physikalisch-Technische Bundesanstalt (PTB) and its implications for their optimal application are provided, as well as information on future challenges and expected progress. Autocollimator-based deflectometric profilometers are in operation at the PTB, the Advanced Light Source (ALS), Lawrence Berkeley National Laboratory, USA, the BESSY II, Helmholtz-Zentrum Berlin fuer Materialien und Energie GmbH, Germany, and the Diamond Light Source Ltd (DLS), UK. A continuous topic of research at the PTB is to investigate the factors influencing the angular response of autocollimators and to provide the most accurate calibration of the instruments to aid users in their application and manufacturers in their improvements [1-3]. At the PTB, the calibration of autocollimators is realized by a direct comparison of the devices with the primary angle reference standard [1]. With highly stable autocollimators, calibrations with standard measurement uncertainties of u = 0.003 arcsec (15 nrad) have been achieved [2,3]. The influence of a number of parameters on the angle measurement with autocollimators have been investigated in detail, such as the reflectivity of the surface under test (SUT), the diameter and shape of the aperture stop, its position both along the autocollimator's optical axis and perpendicular to it [2], as well as the optical path length of the autocollimator beam [3]. Extensive information on these errors and their minimization will be presented.