Search NASA⌕ Search

SEARCH · Search NASA

Results for “Dyson spectrometer”

Search indexed NASA NTRS and DOE OSTI research on propulsion, heat transfer, battery materials and energy systems. Follow report and document links to the original sources.

Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.

Long-Wave Infrared Dyson Spectrometer

Preliminary results are presented for an ultra compact long-wave infrared slit spectrometer based on the dyson concentric design. The dyson spectrometer has been integrated in a dewar environment with a quantum well infrared photodetecor (QWIP), concave electron beam fabricated diffraction grating and ultra precision slit. The entire system is cooled to cryogenic temperatures to maximize signal to noise ratio performance, hence eliminating thermal signal from transmissive elements and internal stray light. All of this is done while maintaining QWIP thermal control. A general description is given of the spectrometer, alignment technique and predicated performance. The spectrometer has been designed for optimal performance with respect to smile and keystone distortion. A spectral calibration is performed with NIST traceable targets. A 2-point non-uniformity correction is performed with a precision blackbody source to provide radiometric accuracy. Preliminary laboratory results show excellent agreement with modeled noise equivalent delta temperature and detector linearity over a broad temperature range.

imaging↗

Radiometric and spectral stray light correction for the Portable Remote Imaging Spectrometer (PRISM) coastal ocean sensor

The airborne Portable Remote Imaging Spectrometer (PRISM) instrument is based on a fast (F/1.8) Dyson spectrometer operating at 350-1050 nm and a two-mirror telescope combined with a Teledyne HyViSI 6604A detector array. Raw PRISM data contain electronic and optical artifacts that must be removed prior to radiometric calibration. We provide an overview of the process transforming raw digital numbers to calibrated radiance values. Electronic panel artifacts are first corrected using empirical relationships developed from laboratory data. The instrument spectral response functions (SRF) are reconstructed using a measurement-based optimization technique. Removal of SRF effects from the data improves retrieval of true spectra, particularly in the typically low-signal near-ultraviolet and near-infrared regions. As a final step, radiometric calibration is performed using corrected measurements of an object of known radiance. Implementation of the complete calibration procedure maximizes data quality in preparation for subsequent processing steps, such as atmospheric removal and spectral signature classification.

Thompson, David R.↗

Europa Environmental Testing of Thermal Hardware for the Mapping Imaging Spectrometer for Europa (MISE) Instrument

The Europa Clipper spacecraft, scheduled for launch in June 2022, hosts a suite of instruments including the Mapping Imaging Spectrometer for Europa (MISE). MISE is a high-optical through-put pushbroom infrared imaging spectrometer that can collect measurements within Europa’s challenging radiation environment. It is externally mounted to the spacecraft and as a result is exposed to high radiation levels and cold temperatures. The instrument consists of a scan mirror assembly, an optical bench including a Dyson spectrometer and a telescope, and a structure that supports a cryocooler and a radiator. The cryocooler is used to actively cool the focal plane array and the spectrometer. The heat from the cryocooler is rejected through pyrolytic graphite sheet thermal straps to a honeycomb radiator. MISE will implement thermal hardware that will either not be used on other parts of the spacecraft or have a more extreme environment than the spacecraft. As a result, the MISE instrument performed numerous environmental tests on its thermal hardware. This paper discusses the results of the thermal cycling and radiation tests performed on the MISE thermal hardware including a Lockheed Martin Micro1-2 cryocooler, Pyrolytic graphite sheet thermal straps, Annealed pyrolytic graphite K-Core, Rosemount Platinum Resistance Thermometers, Tayco Kapton Thermofoil heaters, Dale Ohm resistors, and aluminum honeycomb and facesheets.

Strong, Patrick↗

Optical Design of the Earth Surface Mineral Dust Source Investigation (EMIT) Imaging Spectrometer

The Earth Surface Mineral Dust Source Investigation (EMIT) instrument is a high fidelity imaging spectrometer developed to characterize surface mineralogy of the Earth’s dust source regions over the spectral range of 380- 2500 nm and spectral sampling of 7.4 nm. EMIT will close the current knowledge gap in dust source mineral composition by collecting over 1 billion high signal-to-noise ratio spectra in this region of our planet. These new measurements will be used in conjunction with state-of-the-art Earth System Models to understand and reduce the uncertainty in the radiative forcing effect of mineral dust aerosols. EMIT will be deployed on the International Space Station that has an orbit that is well suited for measuring the arid land regions of the Earth. The optical design utilizes a Dyson spectrometer to reduce volume and mass for a fast (F/1.8) and wide swath (1240 samples) optical system. An overview of the EMIT optical design, development, and current status are discussed.

Green, Robert O.↗

Airborne Visible/Infrared Imaging Spectrometer 3 (AVIRIS-3)

The Airborne Visible/Infrared Imaging Spectrometer 3 (AVIRIS-3) is the third of the NASA AVIRIS spectrometer series and is being developed in parallel with the Compact Wide-swath Imaging Spectrometer II (CWIS-II) for the University of Zurich, Switzerland. The core spectrometer of AVIRIS-3 is a copy of the optically fast, F/1.8 Dyson imaging spectrometer used by the Earth Surface Mineral Dust Source Investigation (EMIT) that is in development and scheduled for launch to the International Space Station (ISS) in 2022. AVIRIS-3 is intended to provide state-of-the-art imaging spectroscopy measurements for NASA science and application through the next decade and beyond. AVIRIS-3 uses the EMIT spectrometer design interfaced with a scaled two mirror telescope enclosed in a portable vacuum vessel to enable measurements from airborne platforms ranging from a Twin Otter to a business jet or a NASA ER-2. AVIRIS-3 is a cryogenic instrument with advanced system control and real-time onboard spectroscopic data processing algorithms evolved from AVIRIS-NG. The spectral range of AVIRIS-3 is 380 to 2500 nm with 7.4 nm sampling. The radiometric range is from 0 to max terrestrial Lambertian radiance with higher signal-to-noise ratio performance than AVIRIS-Classic or AVIRIS-Next Generation. The spatial field-of-view is 39.5 degrees with 0.56 milliradian sampling. This paper describes the design and development of AVIRIS-3 and presents its characteristics in comparison to the previous generation imaging spectrometers.

Helmlinger, Mark↗

Landsat-Swath Imaging Spectrometer Design

We describe the design of a high-throughput pushbroom imaging spectrometer and telescope system that is capable of Landsat swath and resolution while providing better than 10 nm per pixel spectral resolution. The design is based on a 3200 x 480 element x 18 μm pixel size focal plane array, two of which are utilized to cover the full swath. At an optical speed of F/1.8, the system is the fastest proposed to date to our knowledge. The utilization of only two spectrometer modules fed from the same telescope reduces system complexity while providing a solution within achievable detector technology. Predictions of complete system response are shown. Also, it is shown that detailed ghost analysis is a requirement for this type of spectrometer and forms an essential part of a complete design.

Dyson spectrometer↗

A Compact, Fast, Wide-Field Imaging Spectrometer System

We present test results from a compact, fast (F/1.4) imaging spectrometer system with a 33 degree field of view, operating in the 450-1650 nm wavelength region with an extended response InGaAs detector array. The system incorporates a simple two-mirror telescope and a steeply concave bilinear groove diffraction grating made with gray scale x-ray lithography techniques. High degree of spectral and spatial uniformity (97%) is achieved.

diffraction grating↗

Snow and Water Imaging Spectrometer (SWIS): Development of a CubeSat-Compatible Instrument

The Snow and Water Imaging Spectrometer (SWIS) is a fast, high-uniformity, low-polarization sensitivity imaging spectrometer and telescope system designed for integration on a 6U CubeSat platform. Operating in the 350-1700 nm spectral region with 5.7 nm sampling, SWIS is capable of simultaneously addressing the demanding needs of coastal ocean science and snow and ice monitoring. New key technologies that facilitate the development of this instrument include a linear variable anti-reflection (LVAR) detector coating for stray light management, and a single drive on-board calibration mechanism utilizing a transmissive diffuser for solar calibration. We provide an overview of the SWIS instrument design, spacecraft configuration design, and potential science missions.

Dyson spectrometer↗

HyTES: Thermal Imaging Spectrometer Development

The Jet Propulsion Laboratory has developed the Hyperspectral Thermal Emission Spectrometer (HyTES). It is an airborne pushbroom imaging spectrometer based on the Dyson optical configuration. First low altitude test flights are scheduled for later this year. HyTES uses a compact 7.5-12 micrometer m hyperspectral grating spectrometer in combination with a Quantum Well Infrared Photodetector (QWIP) and grating based spectrometer. The Dyson design allows for a very compact and optically fast system (F/1.6). Cooling requirements are minimized due to the single monolithic prism-like grating design. The configuration has the potential to be the optimal science-grade imaging spectroscopy solution for high altitude, lighter-than-air (HAA, LTA) vehicles and unmanned aerial vehicles (UAV) due to its small form factor and relatively low power requirements. The QWIP sensor allows for optimum spatial and spectral uniformity and provides adequate responsivity which allows for near 100mK noise equivalent temperature difference (NEDT) operation across the LWIR passband. The QWIP's repeatability and uniformity will be helpful for data integrity since currently an onboard calibrator is not planned. A calibration will be done before and after eight hour flights to gage any inconsistencies. This has been demonstrated with lab testing. Further test results show adequate NEDT, linearity as well as applicable earth science emissivity target results (Silicates, water) measured in direct sunlight.

remote sensing↗

Compact Infrared Spectrometers

Concentric spectrometer forms are advantageous for constructing a variety of systems spanning the entire visible to infrared range. Spectrometer examples are given, including broadband or high resolution forms. Some issues associated with the Dyson catadioptric type are also discussed.

imaging spectroscopy↗

The Hyperspectral Thermal Emission Spectrometer (HyTES): Preliminary Results

The Hyperspectral Thermal Emission Spectrometer (HyTES) is being developed as part of the risk reduction activities associated with the Hyperspectral Infrared Imager (HyspIRI). HyspIRI is one of the Tier 2 Decadal Survey Missions. HyTES will provide information on how to place the filters on the HyspIRI Thermal Infrared Instrument (TIR) as well as provide antecedent science data. The pushbroom design has 512 spatial pixels over a 50-degree field of view and 256 spectral channels between 7.5 micrometers to 12 micrometers. HyTES includes many key enabling state-of-the-art technologies including a high performance convex diffraction grating, a quantum well infrared photodetector (QWIP) focal plane array, and a compact Dyson-inspired optical design. The Dyson optical design allows for a very compact and optically fast system (F/1.6). It also minimizes cooling requirements due to the fact it has a single monolithic prism-like grating design which allows baffling for stray light suppression. The monolithic configuration eases mechanical tolerancing requirements which are a concern since the complete optical assembly is operated at cryogenic temperatures ((is) approximately 100K). The QWIP allows for optimum spatial and spectral uniformity and provides adequate responsivity or D-star to allow 200mK noise equivalent temperature difference (NEDT) operation across the LWIR passband. Assembly of the system is nearly complete. After completion, alignment results will be presented which show low keystone and smile distortion. This is required to minimize spatial-spectral mixing between adjacent spectral channels and spatial positions. Predictions show the system will have adequate signal to noise for laboratory calibration targets.

Moon Mineralogical Mapper (M3).↗