Calorimetric low temperature detectors for low-energetic heavy ions and their application in accelerator mass spectrometry
The energy-sensitive detection of heavy ions with calorimetric low temperature detectors was investigated in the energy range of E=0.1-1 MeV/amu, commonly used for accelerator mass spectrometry (AMS). The detectors used consist of sapphire absorbers and superconducting aluminum transition edge thermometers operated at T{approx}1.5 K. They were irradiated with various ion beams ({sup 13}C,{sup 197}Au,{sup 238}U) provided by the VERA tandem accelerator in Vienna, Austria. The relative energy resolution obtained was {Delta}E/E=(5-9)x10{sup -3}, even for the heaviest ions such as {sup 238}U. In addition, no evidence for a pulse height defect was observed. This performance allowed for the first time to apply a calorimetric low temperature detector in an AMS experiment. The aim was to precisely determine the isotope ratio of {sup 236}U/{sup 238}U for several samples of natural uranium, {sup 236}U being known as a sensitive monitor for neutron fluxes. Replacing a conventionally used detection system at VERA by the calorimetric detector enabled to substantially reduce background from neighboring isotopes and to increase the detection efficiency. Due to the high sensitivity achieved, a value of {sup 236}U/{sup 238}U=6.1x10{sup -12} could be obtained, representing the smallest {sup 236}U/{sup 238}U ratio measured at the time. In addition, we contributed to establishing an improved material standard of {sup 236}U/{sup 238}U, which can be used as a reference for future AMS measurements.