At least 19 records

Use of an Internal Mercury-Argon Lamp for the Calibration of Raman Frequency in Large Raman Images of Astromaterials

Raman imaging is an emerging tool for the analysis of meteorites, as it is capable of describing mineralogy, carbon composition/speciation, and crystal orientation all within a petrographic context. Adequately examining the heterogeneous composition of a meteorite, while maintaining low laser powers needed to prevent damage to meteorite mineralogy and organics, often requires the collection of large Raman images with long collection times on the timescale of days. However, the frequency calibration of the Raman instrument/spectrometer drifts on long timescales resulting in low frequency precision for the Raman spectra comprising a large Raman image. This decrease in frequency precision can compromise the analysis of geological markers that can provide important information on a meteorite sample. We examine, in detail, the change in bandwidth and frequency of the Raman bands and Hg-Ar emission lines in our Raman images as a function of time and laboratory temperature. To overcome the drift in calibration, we utilize a commercial WITec Raman instrument with an internal Hg-Ar calibration lamp to individually calibrate each spectrum in the Raman image. Our instrument, known as “Ratatoskr”, uses a beam splitter in place of the customary mirror to facilitate collection of Hg-Ar calibration lines concurrent with Raman spectra. We show that using the internal calibration we can improve the frequency precision for the spectra in our Raman images from ~±0.15 cm(exp -1) to ~±0.05 cm(exp -1) for Raman spectra collected over multiple days. This is important as it improves measurements of mineral chemistry, latent strain, and other features that are dependent upon accurate peak position determination. We also examine the spectral signal-to-noise ratio needed to minimize the frequency error when fitting bands to Gaussian/Lorentzian profiles. We then use our results to suggest a general method for calibrating the frequency of Raman spectra in large Raman images.

Jakubek, Ryan S.↗

Use of an Internal Mercury-Argon Lamp for the Calibration of Raman Frequency in Large Raman Images of Astromaterials

Raman imaging is an emerging tool for the analysis of meteorites, as it is capable of describing the mineralogy and organics within a petrographic context. Adequately examining the heterogeneous composition of a meteorite often requires the collection of large Raman images with long collection times on the time-scale of days. However, the frequency calibration of the Raman instrument/spectrometer drifts on long timescales resulting in low frequency precision for the Raman spectra comprising a large Raman image. Here we utilize a Hg-Ar calibration lamp integrated into our Raman microscope to calibrate the Raman spectra within a Raman image. We characterize this method of calibration in detail.

Ryan S. Jakubek↗

Low-frequency Raman signature of Ag-intercalated few-layer MoS 2

Density functional theory based calculations and experimental analysis on a limited number of real samples are performed to study how the presence of silver intercalated in the van der Waals gap of few-layer MoS 2 affects the low-frequency Raman active modes of this material. Silver is found to predominantly affect the breathing-like and shear-like vibrational modes of MoS 2 . These modes correspond to quasi-rigid movements of each individual layer with a restoring force (and, in turn, frequency) that is determined by modulations in the weak interlayer interactions. Noticeable red-shifts with increasing Ag concentration are found for all low-frequency modes. This finding indicates the potential for low-frequency vibrations as useful gauges for practical determination of silver concentration using low-frequency Raman spectroscopy. Finally, this work also describes a semi-classical linear chain model that allows to extrapolate results to a large number of layers. Further, first-principles calculations show how Raman spectroscopy can be used to characterize the quality of the two-dimensional interface between MoS 2 and a silver substrate.

36 MATERIALS SCIENCE↗

Fine Structure of the Low-Frequency Raman Phonon Bands of Single-Wall Carbon Nanotubes

The Raman spectra of singled-wall carbon nanotubes (SWNT) produced by laser and are process were studied between 5 and 500 kappa. The line width vs. temperature dependence of the low-frequency Raman bands between 150 and 200/ cm deviates from that expected for phonon decay through phonon-phonon scattering mechanism. The experimental results and their analysis provided convincing evidence that each of the low-frequency Raman lines is a superposition of several narrower Raman lines corresponding to tubes of nearly the same diameter. The application of Raman spectroscopy to probe the distribution of SWNT by both diameter and chirality is discussed.

Iliev, M. N.↗

Low-frequency Raman active modes of twisted bilayer MoS 2

Abstract We study the low-frequency Raman active modes of twisted bilayer MoS 2 for several twist angles using a force-field approach and a parametrized bond polarizability model. We show that twist angles near high-symmetry stacking configurations exhibit stacking frustration that leads to significant buckling of the moiré superlattice. We find that atomic relaxation due to the twist is of prime importance. The periodic displacement of the Mo atoms shows the realization of a soliton network, and in turn, leads to the emergence of a number of frequency modes not seen in the high-symmetry stacking systems. Some of the modes are only seen in the XZ Raman polarization setup while others are seen in the XY setup. The symmetry of the normal modes, and how this affects the Raman tensors is examined in detail.

75 CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND↗

Low-frequency Raman spectrum of 2D layered perovskites: Local atomistic motion or superlattice modes?

We report the low-frequency Raman spectrum (ω = 10 cm -1 –150 cm -1 ) of a wide variety of alkylammonium iodide based 2D lead halide perovskites (2D LHPs) as a function of A-site cation (MA = methylammonium and FA = formamidinium), octahedral layer thickness (n = 2–4), organic spacer chain length (butyl-, pentyl-, hexyl-), and sample temperature (T = 77 K–293 K). Using density functional theory calculations under the harmonic approximation for n = 2 BA:MAPbI, we assign several longitudinal/transverse optical phonon modes between 30 cm -1 and 100 cm -1 , the eigen displacements of which are analogous to that observed previously for octahedral twists/distortions in bulk MAPbI. Additionally, we propose an alternative assignment for low-frequency modes below this band (<30 cm -1 ) as zone-folded longitudinal acoustic phonons corresponding to the periodicity of the entire layered structure. Furthermore, we compare measured spectra to predictions of the Rytov elastic continuum model for zone-folded dispersion in layered structures. Our results are consistent across the various 2D LHPs studied herein, with energetic shifts of optical phonons corresponding to microscopic structural differences between materials and energetic shifts of acoustic phonons according to changes in the periodicity and elastic properties of the perovskite/organic subphases. This study highlights the importance of both the local atomic order and the superlattice structure on the vibrational properties of layered 2D materials.

2D materials↗

Probing Local Optical Fields via Ultralow Frequency Raman Scattering from a Corrugated Probe

Here we revisit nanoscale local optical field imaging via tip-enhanced Raman scattering (TERS). Rather than taking advantage of molecular reporters to probe different aspects of the local fields, we show how ultralow frequency Raman (ULF) scattering from the (nano-corrugated) metallic probe itself can be used for the same purpose. The bright ULF-TERS response we record allows non-invasive (tapping mode feedback) local field imaging, enables visualizing the local fields of small (down to 20 nm) isolated plasmonic particles, and can also be exploited to distinguish between Si and SiO2 domains with 5 nm spatial resolution. We describe our approach and its limitations, particularly when it comes to using all-metallic vs molecular reporters.

37 INORGANIC, ORGANIC, PHYSICAL, AND ANALYTICAL CH↗

Plume Characterization of a Laboratory Model 22 N GPIM Thruster via High-Frequency Raman Spectroscopy

The Green Propellant Infusion Mission (GPIM) will demonstrate the capability of a green propulsion system, specifically, one using the monopropellant, AF-M315E. One of the risks identified for GPIM is potential contamination of sensitive areas of the spacecraft from the effluents in the plumes of AF-M315E thrusters. Plume characterization of a laboratory-model 22 N thruster via optical diagnostics was conducted at NASA GRC in a space-simulated environment. A high-frequency pulsed laser was coupled with an electron-multiplied ICCD camera to perform Raman spectroscopy in the near-field, low-pressure plume. The Raman data yielded plume constituents and temperatures over a range of thruster chamber pressures and as a function of thruster (catalyst) operating time. Schlieren images of the near-field plume enabled calculation of plume velocities and revealed general plume structure of the otherwise invisible plume. The measured velocities are compared to those predicted by a two-dimensional, kinetic model. Trends in data and numerical results are presented from catalyst mid-life to end-of-life. The results of this investigation were coupled with the Raman and Schlieren data to provide an anchor for plume impingement analysis presented in a companion paper. The results of both analyses will be used to improve understanding of the nature of AF-M315E plumes and their impacts to GPIM and other future missions.

Raman Spectroscopy↗

Selective Detection in Impulsive Low-Frequency Raman Imaging Using Shaped Probe Pulses

Impulsive stimulated Raman scattering (ISRS) using a single short femtosecond pump pulse to excite molecular vibrations offers an elegant pump-probe approach to perform vibrational imaging below 200 cm –1 . One shortcoming of ISRS is its inability to offer vibrational selectivity as all the vibrational bonds whose frequencies lie within the short pump-pulse bandwidth are excited. To date, several coherent control techniques have been explored to address this issue and selectively excite a specific molecular vibration by shaping the pump pulse. There has not been any systematic work that reports an analogous shaping of the probe pulse to implement preferential detection. In this work, we focus on vibrational imaging and report vibrational selective detection by shaping the probe pulse in time. Here, we demonstrate numerically and experimentally two pulse-shaping strategies with one functioning as a vibrational notch filter and the other functioning as a vibrational low-pass filter. This enables fast (25 μs/pixel) and selective hyperspectral imaging in the low-frequency regime (< 200 cm –1 ).

74 ATOMIC AND MOLECULAR PHYSICS↗

Analysis of torsional spectra of molecules with two internal C3v rotors. II - Far infrared and low frequency Raman spectra of dimethylether isotopes

The torsional far infrared and Raman spectra of gaseous CH3OCH3, CD3OCH3, and CD3OCD3 are presented. They are analyzed using a computer program which is based on the results of an extensive investigation of the isometric groups and of the symmetry groups of the rotation-internal rotation Hamiltonians of a series of semirigid two-top models. Four or more Fourier coefficients of the potential functions in two variables could be determined for each isotope. Strong evidence was found for Fermi-resonance-type interactions with the COC bending mode.

Groner, P.↗

Determining the twist angle of stacked MoS 2 layers using machine learning-assisted low-frequency interlayer Raman fingerprints

The investigation of twisted stacked few-layer MoS 2 has revealed novel electronic, optical, and vibrational properties over an extended period. For the successful integration of twisted stacked few-layer MoS 2 into a wide range of applications, it is crucial to employ a noninvasive, versatile technique for characterizing the layered architecture of these complex structures. Here, in this work, we introduce a machine learning-assisted low-frequency Raman spectroscopy method to characterize the twist angle of few-layer stacked MoS 2 samples. A feedforward neural network (FNN) is utilized to analyze the low-frequency breathing mode as a function of the twist angle. Moreover, using finite difference method (FDM) and density functional theory (DFT) calculations, we show that the low-frequency Raman spectra of MoS 2 are mainly influenced by the effect of the nearest and second nearest layers. A new improved linear chain model (TA-LCM) with taking the twist angle into the consideration is developed to understand the interlayer breathing modes of stacked few-layer MoS 2 . This approach can be extended to other 2D materials systems and provides an intelligent way to investigate naturally stacked and twisted interlayer interactions.

37 INORGANIC, ORGANIC, PHYSICAL, AND ANALYTICAL CH↗

Low frequency coherent Raman spectroscopy

Abstract We revisit low frequency coherent Raman spectroscopy (LF-CRS) and present a unified theoretical background that provides consistent physical pictures of LF-CRS signal generation. Our general framework allows to compute the signal to noise ratio in the multitude of possible LF-CRS, and more generally CRS, experimental implementations both in the spectral and time domain.

Bartels, Randy A. (ORCID:0000000305300435)↗

Low Frequency Vibrational Modes of Two-Dimensional Lead-Free Metal Halide Double Perovskites

2D layered double perovskites of (S-MPA) 4 AgBiI 8 (MPA-AgBiI 8 ) and (S-MPA) 4 CuBiI 8 (MPA-CuBiI 8 ) (S-MPA, S-β-methylphenethylammonium) were synthesized with a hydrothermal method. The crystal structure of MPA-AgBiI 8 was determined using single-crystal X-ray diffraction (scXRD). Powder XRD (pXRD) data suggest that the crystal structure of MPA-CuBiI 8 is more complex than that of MPA-AgBiI 8 . UV-Vis electronic absorption spectra of these perovskites reveal a bandgap of 2.03 eV for both. Short exciton lifetime from time-resolved photoluminescence (TRPL) results and low PL intensity of the Cu-based perovskite suggest a high density of trap states within the bandgap. Low frequency Raman spectra of both materials show distinct peaks and a slightly higher frequency for the Cu-based perovskite than the Ag-based perovskite. Density functional theory (DFT) calculations were conducted to simulate the low frequency Raman spectra and help explain the different phonon modes, which are collective vibrations of metal halide bond bending and stretching within the Ag- and Cu-centered octahedra coupled with MPA libration and twisting within the inorganic layer. The DFT theory also quantified octahedral distortions in the two perovskites. Furthermore, this combined experimental and computational study provides new insights into the low frequency vibrations of 2D perovskites.

Halogens↗

Development of a two photon/laser induced fluorescence technique for the detection of atmospheric OH radicals

The development of a new mid-IR laser source was the primary goal. Backward propagating stimulated D2 Raman frequency down conversion of a commercially available 1.06 micron Nd:YAG laser was shown to generate an efficient source of 1.56 micron radiation with near diffraction limited beam quality. The efficient generation of a 2.9 micron laser source was also achieved using backward propagating CH4 Raman frequency down conversion of the 1.56 micron pump. Slightly higher efficiencies were obtained for frequency down conversion of the 1.06 micron Nd:YAG using the H2 Raman shift yielding a near diffraction limited source in the 200 mJ range at 1.9 micron. Similar conversion efficiencies are anticipated as a result of extending the wavelength coverage of recently available Ti:sapphire pulse laser to not only cover the 740 to 860 nm fundamental wavelength range but also the .95 to 1.15 and 1.06 to 1.33 micron range using D2 and H2, respectively. The anticipated sensitivity of a TP-LIF OH sensor using this mid-IR source would give signal limited detection of 1.4 x 10(exp 5) OH/cu cm under boundary layer conditions and 5.5 x 10(exp 4) OH/cu cm under free troposphere sampling conditions for a five minute signal integration period. This level of performance coupled with the techniques non-perturbing nature and freedom from both interferences and background would allow reliable tropospheric OH measurement to be obtained under virtually any ambient condition of current interest, including interstitial and sampling.

Bradshaw, John↗