Analyze the real-time performance of the optimized fiber-PTR instrument
An instrument to perform in-reactor, real-time thermal property measurements of nuclear fuels and materials was recently designed, optimized, and tested at Massachusetts Institute of Technology Research Reactor. It was the second insertion experiments of a total of two rounds. Here we report the collection and analysis of the real-time captured thermal diffusivity data on the preloaded SiC. With the optimized design, the instrument temperature had been successfully raised as we expected. Only one probe (the top one) returned real-time thermal diffusivity in a time span of approximately two days. Measured thermal diffusivity of the preloaded SiC was significantly lower than the literature value at the same temperature, indicating a significant influence from the irradiation induced microstructure damage. Meanwhile, the real-time thermal diffusivity value continuously reduced with time, indicating the dynamic microstructure evolution. The possible reasons of the probes that did not return data were hypothesized. The further optimization of the instrument with the objective to improve the measurement accuracy and instrument survivability is also discussed.