Search NASASearch

SEARCH · Search NASA

Results for “VISCOMETER”

Search indexed NASA NTRS and DOE OSTI research on propulsion, heat transfer, battery materials and energy systems. Follow report and document links to the original sources.

Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.

At least 19 records

Calibration of the Oscillating Screen Viscometer

We have devised a calibration procedure for the oscillating screen viscometer which can provide the accuracy needed for the flight measurement of viscosity near the liquid-vapor critical point of xenon. The procedure, which makes use of the viscometer's wide bandwidth and hydrodynamic similarity, allows the viscometer to be self-calibrating. To demonstrate the validity of this procedure we measured the oscillator's transfer function under a wide variety of conditions. We obtained data using CO2 at temperatures spanning a temperature range of 35 K and densities varying by a factor of 165, thereby encountering viscosity variations as great as 50%. In contrast the flight experiment will be performed over a temperature range of 29 K and at only a single density, and the viscosity is expected to change by less than 40%. The measurements show that, after excluding data above 10 Hz (where frequency-dependent corrections are poorly modeled) and making a plausible adjustment to the viscosity value used at high density, the viscometer's behavior is fully consistent with the use of hydrodynamic similarity for calibration. Achieving this agreement required understanding a 1% anelastic effect present in the oscillator's torsion fiber.

Berg, Robert F.

Viscometer for low frequency, low shear rate measurements

A computer-controlled torsion-oscillator viscometer with low 0.5 Hz frequency and very low 0.05/s shear rate is designed to precisely study shear-sensitive fluids such as microemulsions, gels, polymer solutions and melts, colloidal solutions undergoing coagulation, and liquid mixtures near critical points. The viscosities are obtained from measurements of the logarithmic decrement of an underdriven oscillator. The viscometer is found to have a resolution of 0.2 percent when used with liquid samples and a resolution of 0.4 percent when used with a dense gaseous sample. The design is compatible with submillikelvin temperature control.

Berg, R. F.

NIST torsion oscillator viscometer response: Performance on the LeRC active vibration isolation platform

Critical point viscosity measurements are limited to their reduced temperature approach to T(sub c) in an Earth bound system, because of density gradients imposed by gravity. Therefore, these classes of experiments have been proposed as good candidates for 'microgravity' science experiments where this limitation is not present. The nature of these viscosity measurements dictate hardware that is sensitive to low frequency excitations. Because of the vibratory acceleration sensitivity of a torsion oscillator viscometer, used to acquire such measurements, a vibration isolation sensitivity test was performed on candidate 'microgravity' hardware to study the possibility of meeting the stringent oscillatory sensitivity requirements of a National Institute of Standards and Technology (NIST) torsion oscillator viscometer. A prototype six degree of freedom active magnetic isolation system, developed at NASA Lewis Research Center, was used as the isolation system. The ambient acceleration levels of the platform were reduced to the noise floor levels of its control sensors, about one microgravity in the 0.1 to 10 Hz bandwidth.

Berg, Robert F.

The Shuttle's Vibration Environment is Compatible with the Oscillating Screen Viscometer

The oscillating screen viscometer is designed for useful operation in the vibration environment of the Space Shuttle. The following analysis compares recently analyzed on-orbit vibration data against the viscometer's measured vibration sensitivity. During routine crew activities the effect of the resulting vibration on the precision of the viscosity measurement will be tolerable. However, there is a possible problem if the astronauts use unisolated exercise equipment for extended periods of time.

Berg, Robert F.

Analysis of variable-head viscometer data

Nonnewtonian mathematical model and plotting method for analyzing variable head viscometer data obtained on pseudo-plastic polymer solutions - Rheology

MATHEMATICAL MODEL

Parallel-plate viscometer

Viscometer consists of movable vertical rod with one optical flat fixed to its lower end and centered over second optical flat held rigidly parallel to moveable flat. Two perforated diaphragms of thin metal permit limited amount of vertical movement of rod carrying movable flat, but resist lateral movement.

Fearnehough, H. T.

Rheology melts and magmatic suspensions. I - Design and calibration of concentric cylinder viscometer with application to rhyolitic magma

The design and calibration of concentric cylinder viscometer for rhyolitic magma applications are described together with the methods of data reduction and error analysis. Experimental data are presented on two rhyolitic magmas (melt plus a small fraction of vapor) under conditions of varying temperature (1100-1350 C) and shear rate (0.05-13.0/sec) at 100 kPa total pressure. Data obtained include a first reported measurement of a normal stress coefficient for magma.

Spera, Frank J.

A hydrodynamic model of the oscillating screen viscometer

The viscometer consists of an oscillating screen immersed in a fluid and free to rotate about an axis in its plane. The viscosity can be determined from the measured ratio of the periodic driving force to the screen motion when an adequate hydrodynamical model of the immersed oscillator is available. The screen is formed by a square mesh of thin wire whose dimensions invite comparison with asymptotic results for narrow hollow bodies translating in Stokes flow. These indicate that the closed hole structure of the grid plays an important role in determining its motion. It is shown that this role diminishes as the frequency increases. The computed results, obtained from a system of linear equations, are consistent with experimental values over the appropriate range of frequency.

Davis, A. M. J.

Effect of Bulk Viscosity on the Oscillating Screen Viscometer

Close to the critical temperature, the bulk viscosity of the xenon sample will exceed the shear viscosity by more than a factor of a billion. Nevertheless, the viscometer's low operating frequency ensures that the only significant force on the oscillating screen will be due to the shear viscosity.

Berg, Robert F.

True Shear Parallel Plate Viscometer

This viscometer (which can also be used as a rheometer) is designed for use with liquids over a large temperature range. The device consists of horizontally disposed, similarly sized, parallel plates with a precisely known gap. The lower plate is driven laterally with a motor to apply shear to the liquid in the gap. The upper plate is freely suspended from a double-arm pendulum with a sufficiently long radius to reduce height variations during the swing to negligible levels. A sensitive load cell measures the shear force applied by the liquid to the upper plate. Viscosity is measured by taking the ratio of shear stress to shear rate.

Ethridge, Edwin