[Measurements of "Total Water" and Carbon Dioxide from the NASA WB-57 During Crystal-Face]
An existing closed-path tunable diode laser hygrometer (CLH) was employed for the measurements of total water made during CRYSTAL-FACE. This instrument had flown previously on the NASA DC-8 during the SAGE III Ozone Loss and Validation Experiment (SOLVE) and also on the NCAR C-130 during some local flights designed to test the extent of water vapor interference in carbon dioxide measurements. The instrument was largely unchanged from previous studies, but a new inlet appropriate to the WB-57F wingpod was constructed. In order to minimize the impact on the over-subscribed right wingpod and to achieve good thermal control of the inlet temperature, the CLH inlet was made of carbon-fiber/epoxy composite. Considerable effort was spent to design and build the lightest possible mounting hardware and design relatively low-power inlet heaters. As a result, the instrument and mounting hardware came in below the NASA/JSC-imposed weight cap of 35 lbs. Data were obtained on all test flights during May 2002 and during all but one mission flight in July 2002 (the one lost flight was due to an unplugged instrument power cable). Instrument performance during the test flights was good, but the data are not science- quality, as a variety of tests were performed to optimize the inlet configuration and heating. Data on all mission flights is of high quality, despite some difficulties caused by flying through wet low-altitude air masses and dense anvils, which saturated the instrument response.