Applications of cryogenics in electron microscopy

Description of research and development efforts which resulted in a high-voltage cryoelectron microscope system capable of consistent operation at 1.8 to 4.2 K. Attention is given to the design and operation of superconducting objective lenses providing enhanced resolution during longer exposure times at lower beam intensities (thus reducing radiation damage of specimens). A specific system described combines a closed-cycle superfluid helium refrigerator integrated with a modified 200 kV electron microscope. Consistent resolutions of 8 to 16 A are attained with significantly reduced radiation damage, contamination, and thermal noise in prolonged vibration-free examination of specimens at temperatures from 1.8 to 4.2 K. Applications in specific disciplines are discussed, including membrane ultrastructure, cryobiology, microelectronics, and general superconductivity research.

Fernandez-Moran, H.↗