Delta Method Application for the Correlation of IR Detector Thermal Parasitic Loads with Statistically Accurate Results
The Linear Etalon Imaging Spectral Array (LEISA) is a cryogenic IR detector that is passively cooled below 110K. LEISA is part of the L’Ralph instrument, which is going on the Lucy mission, with planned flyby’s of the Trojan asteroids around 5.5 AU. One of the primary thermal challenges is to successfully quantify each significant parasitic heat flow value with statistical certainty.Simulation of flight-like thermal environments and correlation of thermal models is an important part of this challenge, which can be critical for thermally sensitive systems such as cryogenic detectors that are passively cooled. It is common practice during TVAC to achieve several thermal balance points and then to correlate several design variables in order to achieve the closest match to all balance points tested. This approach lends itself to high levels of uncertainty in the correlated values due to the cumulative effect of absolute temperature uncertainty and differences between the sensors used for the measurement. Accuracy can be improved with the use of more accurate sensors, by calibrating all sensors with respect to each other, and by using Zero-Q methodologies, all of which add cost and complexity to the test.The Delta Method uses the difference in temperature caused by the changes in the boundary conditions to calculate heat flows and/or thermal resistances across temperature differentials. This eliminates the problem of high uncertainties with absolute temperature measurements, and yields statistically accurate results down to the sensor resolution of 1mK at < 100K, by taking the sensor’s calibration errors out of the equation. During testing, this method relies on parametric variations around an initial balance point, by varying the boundary conditions of the different parasitic heat sources. The Delta Method, as well as its application on the LEISA thermal test, will be discussed in detail.