Urine monitoring system failure analysis and operational verification test report
Failure analysis and testing of a prototype urine monitoring system (UMS) are reported. System performance was characterized by a regression formula developed from volume measurement test data. When the volume measurement test data. When the volume measurement data was imputted to the formula, the standard error of the estimate calculated using the regression formula was found to be within 1.524% of the mean of the mass of the input. System repeatability was found to be somewhat dependent upon the residual volume of the system and the evaporation of fluid from the separator. The evaporation rate was determined to be approximately 1cc/minute. The residual volume in the UMS was determined by measuring the concentration of LiCl in the flush water. Observed results indicated residual levels in the range of 9-10ml, however, results obtained during the flushing efficiency test indicated a residual level of approximately 20ml. It is recommended that the phase separator pumpout time be extended or the design modified to minimize the residual level.