Modeling Background Attenuation by Sample Matrix in Gamma Spectrometric Analyses
In laboratory gamma spectrometric analyses, the procedures for estimating background usually overestimate it. If an empty container similar to that used to hold samples is measured, it does not consider the background attenuation by sample matrix. If a 'blank' sample is measured, the hypothesis that this sample will be free of radionuclides is generally not true. The activity of this 'blank' sample is frequently sufficient to mask or to overwhelm the effect of attenuation so that the background remains overestimated. In order to overcome this problem, a model was developed to obtain the attenuated background from the spectrum acquired with the empty container. Beyond reasonable hypotheses, the model presumes the knowledge of the linear attenuation coefficient of the samples and its dependence on photon energy and samples densities. An evaluation of the effects of this model on the Lowest Limit of Detection (LLD) is presented for geological samples placed in cylindrical containers that completely cover the top of an HPGe detector that has a 66% relative efficiency. The results are presented for energies in the range of 63 to 2614keV, for sample densities varying from 1.5 to 2.5 g{center_dot}cm{sup -3}, and for the height of the material on the detector of 2 cm and 5 cm. For a sample density of 2.0 g{center_dot}cm{sup -3} and with a 2cm height, the method allowed for a lowering of 3.4% of the LLD for the energy of 1460keV, from {sup 40}K, 3.9% for the energy of 911keV from {sup 228}Ac, 4.5% for the energy of 609keV from {sup 214}Bi, and8.3% for the energy of 92keV from {sup 234}Th. For a sample density of 1.75 g{center_dot}cm{sup -3} and a 5cm height, the method indicates a lowering of 6.5%, 7.4%, 8.3% and 12.9% of the LLD for the same respective energies.