High sensitivity real-time NVR monitor

A real time non-volatile residue (NVR) monitor, which utilizes surface acoustic wave (SAW) resonators to detect molecular contamination in a given environment. The SAW resonators operate at a resonant frequency of approximately 200 MHz-2,000 MHz which enables the NVR monitor to detect molecular contamination on the order of 10.sup.-11 g-cm.sup.-2 to 10.sup.-13 g-cm.sup.2. The NVR monitor utilizes active temperature control of (SAW) resonators to achieve a stable resonant frequency. The temperature control system of the NVR monitor is able to directly heat and cool the SAW resonators utilizing a thermoelectric element to maintain the resonators at a present temperature independent of the environmental conditions. In order to enable the direct heating and cooling of the SAW resonators, the SAW resonators are operatively mounted to a heat sink. In one embodiment, the heat sink is located in between the SAW resonators and an electronic circuit board which contains at least a portion of the SAW control electronics. The electrical leads of the SAW resonators are connected through the heat sink to the circuit board via an electronic path which prevents inaccurate frequency measurement.

Bowers, William D.↗

A Real-Time Nonvolatile Residue (NVR) Monitor

New development and application of device described in "Surface-Acoustic-Wave Piezoelectric Microbalance," (LAR-14476). Active sensing element of Real-Time NVR Monitor comprises pair of piezoelectric surface-acoustic-wave resonators resonating at frequency of 200 MHz. Bare, uncoated resonator exposed to atmosphere and directly in contact with airborne volatile and nonvolatile materials leaving residues on surface. Resonant frequency of exposed resonator decreases with increasing mass of adsorbed residue; resulting beat frequency between two resonators increases with mass and serves as sensitive real-time indication of airborne contaminants or non-volatile residue.

Bowers, William D.↗

Surface-Acoustic-Wave Piezoelectric Microbalance

Improved piezoelectric microbalances developed for use in measuring masses of volcanic, aerosol, and other small particles suspended in air. Sensitive microbalance used to analyze airborne particles in real time in environments as diverse as clean rooms or upper atmosphere. Surface-acoustic-wave resonator includes input and output sets of interdigitated electrodes and two passive conductive patterns acting as reflectors. Mechanical energy travels both ways out from middle and reflected back toward middle. Microbalance and associated circuitry fit in small package. Circuit draws only 80 mA at 5 V. Sensitivity more than 400 times that of bulk piezoelectric microbalance.

Chuan, Raymond L.↗

A high sensitivity real-time NVR monitor

The use of a temperature-controlled 200-MHz SAW resonator piezoelectric mass microbalance to monitor the mass of nonvolatile residue (NVR) deposited on its surface in real time is reported. The fundamental frequency of this device is mainly dependent on the configuration of the transducers and not on the thickness of the substrate. Therefore, higher operating frequencies can be achieved without reducing the thickness of the crystal. The real-time instrument was integrated onto a conventional stainless steel NVR plate and operated flawlessly over a 14-d period at Kennedy Space Center and successfully measured less than 1 ng/sq cm d NVR contamination. Contamination episodes detected by the instrument were correlated with scheduled activities on the test stand. Under the assumption of a baseline noise level of +/- 2 Hz, the absolute mass lower limit of detection would be 0.065 ng/sq cm. This would enable the detection of a daily NVR deposition rate of less than 0.1 ng/sq cm d.

Bowers, William D.↗

A non-optical real-time particle fallout monitor

The paper describes a size-selective fallout monitor that can be employed to assess the degradation of sensitive optical surfaces as well as facilitate the analysis of particle types. The device combines a vertical elutriator and a quartz crystal microbalance, and only particles greater than a specific size can pass through an upward laminar flow generated in the device. The larger particles cause a frequency shift in the crystal oscillator, thereby permitting the measurement of the fallout associated with the contamination of optical instruments.

Chuan, Raymond L.↗

A 200 MHz surface acoustic wave mass microbalance

The principle of operation of the surface acoustic wave (SAW) piezoelectric crystals used as microgravimetric sensors in mass microbalances is discussed. Special attention is given to a SAW 200-MHz crystal developed for measuring molecular deposition on spacecrafts, whose operating frequency does not depend on the thickness of the crystal. The frequency stability of the 200 MHz SAW device is better than 5 x 10 exp -9, which corresponds to a lower limit-of-detection of 3 x 10 exp -12 g for a signal-to-noise ratio of 3. A block diagram of the 200 MHz SAW mass microbalance and a schematic diagram of SAW resonator are presented together with performance data of this device.

Bowers, William D.↗