Content, David

Publications and source records attributed to Content, David.

At least 19 records

Alignment and Test of the Wide Field Infrared Survey Telescope (WFIRST) Engineering Design Unit (EDU) Grism

The WFIRST wide field instrument (WFI) includes a slitless spectrometer, which plays an important role in the WFIRST mission for the survey of emission-line galaxies. WFI is building engineering design and environmental test (EDU and ETU) units to reduce risk for the flight grism unit. We report here on successful build and test of the EDU grism. The four-element EDU grism consists of two prism elements and two diffractive elements that provide R700 dispersion. The elements were fabricated with alignment fiducials and integral flats to allow opto-mechanical alignment in six-degrees of freedom. Each element in turn, was installed onto a hexapod and positioned to its nominal orientation relative to the grism deck, then bonded into individual cells. Alignment measurements were performed in situ using theodolites to set tip/tilt and a Micro-vu non-contact Multisensor Measurement System was used to set despace, decenter and clocking of each element using the hexapod. After opto-mechanical alignment, the grism wavefront was measured using an Infrared ZYGO interferometer at various field points extending over a 20 by 14- degree (local) field of view. Using modeled alignment sensitivities, we determined the alignment correction required on our Element 2 prism compensator and successfully minimized the field dependent wavefront error and confocality. This paper details the alignment and testing of the EDU grism at ambient and cold operating temperatures.

Hagopian, John

Optical Design and Predicted Performance of the WFIRST Phase-B Imaging Optics Assembly and Wide Field Instrument

Now in Phase-B, the architecture of the Wide-Field Infra-Red Survey Telescope (WFIRST) payload has matured since 2013 to accommodate various opto-mechanical constraints. Based on a 2.4-meter aperture Forward Optical Assembly (FOA), the Imaging Optics Assembly (IOA) provides corrected optical fields to each on-board instrument. Using a Three Mirror Anastigmat (TMA) optical design, the Wide-Field Channel (WFC) provides ~1/3-square degree of instantaneous field coverage at 0.11 arcsecond pixel scale. The WFC as-built predictive analysis anticipates near diffraction-limited imaging over a focal plane of 300.8 million pixels, operating in seven panchromatic bands between 0.48 - 2.0μm, or a 1-octive multi-spectral imaging mode from ~0.95-1.93μm. The IOA provides the Coronagraph Instrument (CGI) a collimated beam with very specific wavefront constraints. We present configuration changes since 2013 that improved interfaces, improved testability, and reduced technical risk. We provide an overview of our Integrated Modeling results, performed at an unprecedented level for a phase-A study, to illustrate performance margins with respect to static wavefront error, jitter, and thermal drift.

Dark Energy

Wide-field Infrared Survey Telescope (WFIRST): Composite Structure Verification for Operational Temperatures

The Wide-Field Infrared Survey Telescope (WFIRST) mission is the next large astrophysics observatory for NASA after the James Webb Space Telescope and is the top priority mission from the 2010 National Academy of Sciences' decadal survey. The WFIRST Optical Telescope Assembly (OTA) includes inherited composite support structures that were originally designed and tested for room temperature operation; however, the WFIRST mission will require operation at colder temperatures to achieve sufficient sensitivity for the infrared wavelengths. We will present the results and conclusions of testing completed at the coupon and engineering model level to verify that the inherited composite structures will maintain mechanical integrity and performance over the required temperature range. The testing included: (1) characterization testing of constituent material coupons, (2) thermal cycling and static load testing of a representative aft metering structure (AMS) and forward metering structure (FMS), and (3) thermal cycling and dynamic testing of a representative secondary mirror assembly (SMA).

metering

Wide-Field InfraRed Survey Telescope (WFIRST) Slitless Spectrometer: Design, Prototype, and Results

The slitless spectrometer plays an important role in the Wide-Field InfraRed Survey Telescope (WFIRST) mission for the survey of emission-line galaxies. This will be an unprecedented very wide field, HST quality 3D survey of emission line galaxies. The concept of the compound grism as a slitless spectrometer has been presented previously. The presentation briefly discusses the challenges and solutions of the optical design, and recent specification updates, as well as a brief comparison between the prototype and the latest design. However, the emphasis of this paper is the progress of the grism prototype: the fabrication and test of the complicated diffractive optical elements and powered prism, as well as grism assembly alignment and testing. Especially how to use different tools and methods, such as IR phase shift and wavelength shift interferometry, to complete the element and assembly tests. The paper also presents very encouraging results from recent element tests to assembly tests. Finally we briefly touch the path forward plan to test the spectral characteristic, such as spectral resolution and response.

Grism

Detector Control and Data Acquisition for the Wide-Field Infrared Survey Telescope (WFIRST) with a Custom ASIC

The Wide-Field Infrared Survey Telescope (WFIRST) will have the largest near-IR focal plane ever flown by NASA, a total of 18 4K x 4K devices. The project has adopted a system-level approach to detector control and data acquisition where 1) control and processing intelligence is pushed into components closer to the detector to maximize signal integrity, 2) functions are performed at the highest allowable temperatures, and 3) the electronics are designed to ensure that the intrinsic detector noise is the limiting factor for system performance. For WFIRST, the detector arrays operate at 90 to 100 K, the detector control and data acquisition functions are performed by a custom ASIC at 150 to 180 K, and the main data processing electronics are at the ambient temperature of the spacecraft, notionally approx.300 K. The new ASIC is the main interface between the cryogenic detectors and the warm instrument electronics. Its single-chip design provides basic clocking for most types of hybrid detectors with CMOS ROICs. It includes a flexible but simple-to-program sequencer, with the option of microprocessor control for more elaborate readout schemes that may be data-dependent. All analog biases, digital clocks, and analog-to-digital conversion functions are incorporated and are connected to the nearby detectors with a short cable that can provide thermal isolation. The interface to the warm electronics is simple and robust through multiple LVDS channels. It also includes features that support parallel operation of multiple ASICs to control detectors that may have more capability or requirements than can be supported by a single chip.

data acquisition

WFIRST-AFTA Overview Technology needs summary Mirror Technology Conference 2015

Presentation covers the overview of the science and hardware of the WFIRST-AFTA (Wide-Field Infrared Survey Telescope) (Astrophysics Focused Telescope Assets) mission. It includes an overview of the technology, with an emphasis on optics technology. It also introduces the WFIRST talks that come later, one on the Wide Field Instrument filters and the other on the CoronaGraph Instrument.

Infared Telescopes

Design, Fabrication and Test of WFIRST-AFTA GRISM Assembly

The Wide-Field Infra-Red Survey Telescope (WFIRST) is designed to perform wide-field imaging and slitless spectroscopic survey of the sky. A compound Grism assembly is selected as its slitless dispersing element. Compared to the grisms in several instruments installed on the Hubble space telescope, the challenge of this grism is the much wider Field Of View (FOV), larger dispersion, and smaller f. The image quality of the slitless spectrometer is directly related to how deep of the sky the spectrometer is able to see. This paper discusses how to use a diffractive lens surface to compensate the wavelength dependent aberration created by the grating in non-collimated space to make the spectrometer diffraction limited, as well as how to intelligently combine the two diffractive surfaces (lens and grating) to reduce the unwanted diffraction orders from the grating. The paper also discusses how to improve the diffraction efficiency of the diffractive surfaces, as well as how to align and assemble the optics.

Diffractive Optics

Cryogenic Refractive Index and Coefficient of Thermal Expansion for the S-TIH1 Glass

Using the CHARMS facility at NASA GSFC, we have measured the cryogenic refractive index of the Ohara S-TIH1 glass from 0.40 to 2.53 micrometers and from 120 to 300 K. We have also examined the spectral dispersion and thermo-optic coefficients (dn/dT). We also derived temperature-dependent Sellmeier models from which refractive index may be calculated for any wavelength and temperature within the stated ranges of each model. The S-TIH1 glass we tested exhibited unusual behavior in the thermo-optic coefficient. We found that for delta < 0.5 micrometers, the index of refraction decrease with a decrease in temperature (positive dn/dT). However, the situation was reversed for delta larger than 0.63 micrometers, where the index will increase with a decrease in temperature (negative dn/dT). We also measured the coefficient of thermal expansion (CTE) for the similar batch of S-TIH1 glass in order to understand its thermal properties. The CTE showed a monotonic change with a decrease in temperature.

Quijada, Manuel A.

Dual-Channel Multi-Purpose Telescope

A dual-channel telescope allows for a wide-field telescope design wit h a good, narrow field channel of fewer surfaces for shorter-wavelen gth or planet-finding applications. The design starts with a Korsch three-mirror-anastigmat (TMA) telescope that meets the mission criter ia for image quality over a wide field of view. The internal image a t the Cassegrain focus is typically blurry due to the aberration bala ncing among the three mirrors. The Cassegrain focus is then re-optim ized on the axis of the system where the narrow field channel instru ment is picked off by bending the primary mirror.

Howard, Joseph M.

Gratings Fabricated on Flat Surfaces and Reproduced on Non-Flat Substrates

A method has been developed for fabricating gratings on flat substrates, and then reproducing the groove pattern on a curved (concave or convex) substrate and a corresponding grating device. First, surface relief diffraction grating grooves are formed on flat substrates. For example, they may be fabricated using photolithography and reactive ion etching, maskless lithography, holography, or mechanical ruling. Then, an imprint of the grating is made on a deformable substrate, such as plastic, polymer, or other materials using thermoforming, hot or cold embossing, or other methods. Interim stamps using electroforming, or other methods, may be produced for the imprinting process or if the same polarity of the grating image is required. The imprinted, deformable substrate is then attached to a curved, rigid substrate using epoxy or other suitable adhesives. The imprinted surface is facing away from the curved rigid substrate. As an alternative fabrication method, after grating is imprinted on the deformable substrate as described above, the grating may be coated with thin conformal conductive layer (for example, using vacuum deposition of gold). Then the membrane may be mounted over an opening in a pressured vessel in a manner of a membrane on a drum, grating side out. The pressure inside of the vessel may be changed with respect to the ambient pressure to produce concave or convex membrane surface. The shape of the opening may control the type of the surface curvature (for example, a circular opening would create spherical surface, oval opening would create toroidal surface, etc.). After that, well-known electroforming methods may be used to create a replica of the grating on the concave or convex membrane. For example, the pressure vessel assembly may be submerged into an electro-forming solution and negative electric potential applied to the metal coated membrane using an insulated wire. Positive electric potential may be then applied to a nickel or other metal plate submerged into the same solution. Metal ions would transfer from the plate through the solution into the membrane, producing high fidelity metal replica of the grating on the membrane. In one variation, an adhesive may be deposited on the deformable substrate, and then cured without touching the rigid, curved substrate. Edges of the deformable substrate may be attached to the rigid substrate to ensure uniform deformation of the deformable substrate. The assembly may be performed in vacuum, and then taken out to atmospheric pressure conditions to ensure that no air is trapped between the deformable and rigid substrates. Alternatively, a rigid surface with complementary curvature to the rigid substrate may be used to ensure uniform adhesion of the deformable substrate to the rigid substrate. Liquid may be applied to the surface of the deformable substrate to uniformly distribute pressure across its surface during the curing or hardening of the adhesive, or the film may be pressed into the surface using a deformable object or surface. After the attachment is complete, the grooves may be coated with reflective or dielectric layers to improve diffraction efficiency.

Content, David

Measuring a Precise Ultra-Lightweight Spaceflight Mirror on Earth: The Analysis of the SHARPI PM Mirror Figure Data during Mirror Processing at GSFC

The SHARPI (Solar High Angular Resolution Photometric Imager) primary mirror is a 5kg, 0.5m paraboloid, diffraction limited at FUV wavelengths when placed in a 0-G environment. The ULE sandwich honeycomb mirror and the attached mount pads were delivered by ITT (then Kodak) in 2003 to NASA s Goddard Space Flight Center (GSFC). At GSFC, we accepted, coated, mounted, and vibration tested this mirror in preparation for flight on the PICTURES (Planet Imaging Concept Testbed Using a Rocket Experiment) mission. At each step, the integrated analysis of interferometer data and FEA models was essential to quantify the 0-G mirror figure. This task required separating nanometer sized variations from hundreds of nanometers of gravity induced distortion. The ability to isolate such features allowed in-situ monitoring of mirror figure, diagnosis of perturbations, and remediation of process errors. In this paper, we describe the technical approach used to achieve these measurements and overcome the various difficulties maintaining UV diffraction-limited performance with this aggressively lightweighted mirror.

Antonille, Scott

Cryogenic System for Interferometry of High-precision Optics at 20 K: Design and Performance

This report describes the facility and experimental methods at the Goddard Space Flight Center Optics Branch for the measurement of the surface figure of cryogenically-cooled spherical mirrors using standard phase-shifting interferometry, with a standard uncertainty below 2nm rms. Two developmental silicon carbide mirrors were tested: both were spheres with radius of curvature of 600 mm, and clear apertures of 150 mm. The mirrors were cooled within a cryostat, and the surface figure error measured through a fused-silica window. The GSFC team developed methods to measure the in-situ SFE at 20 K with a combined standard uncertainty below 2 nm rms and the change in SFE with temperature (the cryochange) with a combined standard uncertainty of 0.25 -- 0.75 nm rms. This paper will present the measurement facility, methods, and uncertainty analysis.

Blake,Peter

Minimizing Uncertainty in Cryogenic Surface Figure Measurement

A new facility at the Goddard Space Flight Center is designed to measure with unusual accuracy the surface figure of mirrors at cryogenic temperatures down to 12 K. The facility is currently configured for spherical mirrors with a radius of curvature (ROC) of 600 mm, and apertures of about 150 mm or less. The goals of the current experiment were to 1) Obtain the best possible estimate of test mirror surface figure, S(x,y) at 87 K and 20 K; 2) Obtain the best estimate of the cryo-change, Delta (x,y): the change in surface figure between room temperature and the two cryo-temperatures; and 3) Determine the uncertainty of these measurements, using the definitions and guidelines of the ISO Guide to the Expression of Uncertainty in Measurement. A silicon mirror was tested, and the cry-change from room temperature to 20K was found to be 3.7 nm rms, with a standard uncertainty of 0.23 nm in the rms statistic. Both the cryo-change figure and the uncertainty are among the lowest such figures yet published. This report describes the facilities, experimental methods, and uncertainty analysis of the measurements.

Blake, Peter

Surface Figure Measurement of Silicon Carbide Mirrors at Cryogenic Temperatures

The surface figure of a developmental silicon carbide mirror, cooled to 87 K and then 20 K within a cryostat, was measured with unusually high precision at the Goddard Space Flight Center (GSFC). The concave spherical mirror, with a radius of 600 mm and a clear aperture of 150 mm, was fabricated of sintered silicon carbide. The mirror was mounted to an interface plate representative of an optical bench, made of the material Cesic@, a composite of silicon, carbon, and silicon carbide. The change in optical surface figure as the mirror and interface plate cooled from room temperature to 20 K was 3.7 nm rms, with a standard uncertainty of 0.23 nm in the rms statistic. Both the cryo-change figure and the uncertainty are among the lowest such figures yet published. This report describes the facilities, experimental methods, and uncertainty analysis of the measurements.

Blake, Peter

Surface Figure Measurement at 20K: Silicon Carbide and Cesic Mirrors

This report presents the facility, methods, and results of testing cryogenically-cooled spherical mirrors, using standard phase-shifting interferometry, at the Goddard Space Flight Center (GSFC). Two mirrors were supplied to GSFC by the European Space Technology Center, so that GSFC could render a second, independent cryo-measurement of their surface figures at 20K. These mirrors, produced by Galileo Avionica and its partners, demonstrate the technology of silicon carbide and SiC-composite lightweighted mirrors, designed for high accuracy (10 nm rms surface figure error) at both room temperature and 20K. The two mirrors provided for testing at GSFC include one made of sintered silicon carbide (mirror blank by Bettini), and a second made of the C/Si/SiC composite known as Cesic (supplied by ECM). Both mirrors are f/2 spheres with 150 mm clear aperture, and with integral mounts. At GSFC, the mirrors will be measured first at at room temperature, making use of standard techniques, with a predicted uncertainty of 2 nm. Then each mirror will be tested in the cryostat, down to 20K: first freely supported and, second, mounted to a Cesic plate that models a Cesic optical bench. The uncertainty of the resultant surface figure error is predicted to be about 5 nm rms. Details of the uncertainty budget are discussed in a related poster. GSFC's techniques and results can be usefully compared to the measurements performed on the same two mirrors by and for Galileo Avionica, who used a somewhat different test configuration and alignment approach.

Blake, Peter