X-ray testing of the AXAF Technology Mirror Assembly (TMA) mirror

The X-ray performance of the Advanced X-ray Astrophysics Facility TMA in the 0.277-6.4-keV range. The precise figure and smooth surface originally specified (which were expected from the in-process optical and mechanical measurements) are verified. The paper primarily describes the test equipment and methods used and gives examples of the detailed on-axis and off-axis data.

Schwartz, D. A.↗

Correspondence between AXAF TMA X-ray performance and models based upon mechanical and visible light measurements

The AXAF Technology Mirror Assembly (TMA) was characterized prior to X-ray testing by properties measured mechanically or with visible light; these include alignment offsets, roundness and global-axial-slope errors, axial-figure errors with characteristic lengths greater than about five mm, and surface roughness with scale lengths between about 0.005 and 0.5 mm. The X-ray data of Schwartz et al. (1985) are compared with predictions based upon the mechanical and visible light measurements.