A dual charge-coupled device /CCD/, astronomical spectrometer and direct imaging camera. I - Optical and detector systems
The MASCOT (MIT Astronomical Spectrometer/Camera for Optical Telescopes), an instrument capable of simultaneously performing both direct imaging and spectrometry of faint objects, is examined. An optical layout is given of the instrument which uses two CCD's mounted on the same temperature regulated detector block. Two sources of noise on the signal are discussed: (1) the CCD readout noise, which results in a constant uncertainty in the number of electrons collected from each pixel; and (2) the photon counting noise. The sensitivity of the device is limited by the sky brightness, the overall quantum efficiency, the resolution, and the readout noise of the CCD. Therefore, total system efficiency is calculated at about 15%.