Abstract
The metabolic analyzer was designed to support experiment M171. It operates on the so-called open circuit method to measure a subject's metabolic activity in terms of oxygen consumed, carbon dioxide produced, minute volume, respiratory exchange ratio, and tidal volume or vital capacity. The system operates in either of two modes. (1) In Mode I, inhaled respiratory volumes are actually measured by a piston spirometer. (2) In Mode II, inhaled volumes are calculated from the exhaled volume and the measured inhaled and exhaled nitrogen concentrations. This second mode was the prime mode for Skylab. Following is a brief description of the various subsystems and their operation.