Calibration analysis for a multi-channel infrared scanning radiometer

Abstract

A procedure for calibrating an infrared scanning spectroradiometer by a computerized parametric error analysis technique was developed. The uncertainties in the radiometric measurements of scene radiance and (for the case of a blackbody scene) temperature due to possible uncertainties in the calibration target temperature, calibration target emissivity, and instrument temperature were calculated for a range of uncertainty levels in the parameters, as well as for a gamut of scent temperatures corresponding to a given spectral channel. This technique is applicable to the radiometric calibration of any infrared radiometer. It was applied specifically to the Cloud-Top Scanning (C.T.S) Radiometer, a three-channel instrument designed for aircraftborne cloud radiance measurements in the 6.75 and 11.5 micron thermal emission spectral regions.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Walden, H., Hurley, E. J., Korb, C. L.. 1977-12-01. Calibration analysis for a multi-channel infrared scanning radiometer. https://ntrs.nasa.gov/citations/19780015960

Cite the original work for its findings. Save a collection to share your selection of sources.