Water vapor measurement system in global atmospheric sampling program, appendix

Abstract

The water vapor measurement system used in the NASA Global Atmospheric Sampling Program (GASP) is described. The system used a modified version of a commercially available dew/frostpoint hygrometer with a thermoelectrically cooled mirror sensor. The modifications extended the range of the hygrometer to enable air sample measurements with frostpoint temperatures down to -80 C at altitudes of 6 to 13 km. Other modifications were made to permit automatic, unattended operation in an aircraft environment. This report described the hygrometer, its integration with the GASP system, its calibration, and operational aspects including measurement errors. The estimated uncertainty of the dew/frostpoint measurements was + or - 1.7 Celsius.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Englund, D. R., Dudzinski, T. J.. 1982-11-01. Water vapor measurement system in global atmospheric sampling program, appendix. https://ntrs.nasa.gov/citations/19830005372