Abstract
This paper presents a characterization of a commercially available position-sensitive detector of energetic ions and neutrals. The detector consists of two microchannel plates followed by a resistive position-encoding anode. The work includes measurement of absolute efficiencies of H(+), He(+), and O(+) ions in the energy range between 250 and 5000 eV, measurement of relative detection efficiencies as a function of particle impact angle, and a simple method for accurate measurement of the time at which a particle strikes the detector.
Keep this discovery
Explore connections, maps & timelines
Gao, R. S., Gibner, P. S., Newman, J. H., Smith, K. A., Stebbings, R. F.. 1984-11-01. Absolute and angular efficiencies of a microchannel-plate position-sensitive detector. https://ntrs.nasa.gov/citations/19850034375