Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) onboard calibration system

Abstract

The AVIRIS instrument uses an onboard calibration system to provide auxiliary calibration data. The system consist of a tungsten halogen cycle lamp imaged onto a fiber bundle through an eight position filter wheel. The fiber bundle illuminates the back side of the foreoptics shutter during a pre-run and post-run calibration sequence. The filter wheel contains two neutral density filters, five spectral filters and one blocked position. This paper reviews the general workings of the onboard calibrator system and discusses recent modifications.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Chrien, Thomas G., Eastwood, Mike, Green, Robert O., Sarture, Charles, Johnson, Howell, Chovit, Chris, Hajek, Pavel. 1995-01-23. Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) onboard calibration system. https://ntrs.nasa.gov/citations/19950027318