Time-Encoded Gamma-Ray Imaging Using a 3-D Position-Sensitive CdZnTe Detector Array
Abstract
The gamma-ray response of thick pixelated CdZnTe detectors varies spatially but is relatively constant as a function of time. Detector irregularities arise from flaws in the material, electrode fabrication, and gaps between the detectors in an array. These nonuniformities generate artifacts in coded aperture images that can obscure a source of interest. This article presents the theory, design, and demonstration of a 3-D CdZnTe-based time-encoded imager. Experimental results show the gamma-ray images of point-like and extended 57 Co and HEU-metal sources formed using a spatially nonuniform 3-D CdZnTe detector array and moving coded aperture mask. Image signal-to-noise ratio (SNR) for a point source was within 10% of its statistical limit at 250k counts.