Focusing ions at atmospheric pressure using nonlinear DC voltage sequences applied to a stacked ring ion guide

Abstract

Many modern ion mobility (IM) and mass spectrometers (MS) operate under low pressure (= 10 Torr) and employ high voltage radiofrequencies (RF) to provide ion confinement. Unfortunately, RF effectiveness drastically decreases as pressure increases, and few techniques for focusing ions at elevated pressures exist. Here we demonstrate a new approach for focusing ions at atmospheric pressure (AP) by applying nonlinear DC voltage sequences following quadratic and power (exponential) functions to a stacked ring ion guide. In this work, we used ion trajectory simulations to rigorously explore how ions react to nonlinear electric fields and validate the simulations with a set of ion current measurements performed at AP. Ion trajectory simulations show that ions initially defocus near the entrance of the device but then become intensely focused as they travel through the device. Contour plots for both nonlinear voltage sequences show electric field lines that increasingly curve inwards as a function of distance, resulting in spatial ion focusing. Experimental ion current and spot size measurements were performed at AP using a 10-cm stacked ring ion guide and a segmented Faraday cup detector. Quadratic sequences produced ~5% smaller spot sizes (~22.8 mm) and ~25% higher ion current compared to a linear voltage sequence (~24.0 mm). Alternatively, power sequences produced ~64% smaller spot sizes (~8.7 mm), albeit with ~10x lower ion current. However, both nonlinear voltage sequences produced similar ion currents at the center of the Faraday cup detector, indicating that higher ion densities are achieved when using nonlinear voltage gradients. These results demonstrate a new way to focus ions at AP, and the capabilities demonstrated here provide fundamental insights on how to keep ions inside analytical devices at elevated pressures without RF.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hollerbach, Adam L., Ibrahim, Yehia M., Norheim, Randolph V., Metz, Thomas O., Ewing, Robert G.. 2022-09-24. Focusing ions at atmospheric pressure using nonlinear DC voltage sequences applied to a stacked ring ion guide. https://doi.org/10.1016/j.ijms.2022.116948

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗