Low-dose high-resolution $\mathrm{TOF}$-$\mathrm{PET}$ using ionization-activated multi-state low-$\mathrm{Z}$ detector media

Abstract

We propose PET scanners using low atomic number media that undergo a persistent local change of state along the paths of the Compton recoil electrons. Measurement of the individual scattering locations and angles, deposited energies, and recoil electron directions allows using the kinematical constraints of the 2-body Compton scattering process to perform a statistical time-ordering of the scatterings, with a high probability of precisely identifying where the gamma first interacted in the detector. In these cases the Line-of-Response is measured with high resolution, determined by the underlying physics processes and not the detector segmentation. There are multiple such media that act through different mechanisms. As an example in which the change of state is quantum-mechanical through a change in molecular configuration, rather than thermodynamic, as in a bubble chamber, we present simulations of a two-state photoswitchable organic dye, a ‘Switchillator’, that is activated to a fluorescent-capable state by the ionization of the recoil electrons. The activated state is persistent, and can be optically excited multiple times to image individual activated molecules. Energy resolution is provided by counting the activated molecules. Location along the LOR is implemented by large-area time-of-flight MCP-PMT photodetectors with single photon time resolution in the tens of ps and sub-mm spatial resolution. Simulations indicate a large reduction of dose.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Shida, J. F., Spieglan, E., Adams, B. W., Angelico, E., Domurat-Sousa, K., Elagin, A., Frisch, H. J., La Riviere, P., Squires, A. H.. 2021-09-14. Low-dose high-resolution $\mathrm{TOF}$-$\mathrm{PET}$ using ionization-activated multi-state low-$\mathrm{Z}$ detector media. https://doi.org/10.1016/j.nima.2021.165801

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Performance evaluation of neutron noise analysis in detecting special nuclear materials

Reliable and rapid inspection techniques play a vital role in preventing illicit trafficking of special nuclear materials. Active interrogation systems using neutrons produced by portable, high-flux deuterium-deuterium or deuterium-tritium neutron generators are being actively developed as a secondary scanning tool for this purpose. In this study, a neutron noise analysis-based approach for detecting unshielded and shielded special nuclear materials by using a pulsed deuterium-tritium neutron generator was evaluated. Here, this approach analyzes the fluctuation of neutron counts. Its performance was quantified with regard to time-to-detection to achieve a minimum probability of detection of 99% and a probability of false alarm of less than 1% considering various amounts of special nuclear materials and different shielding configurations. It was demonstrated that this approach could detect 17 uranium slugs in 5 s given a neutron generator yield of 8.1 × 10 7 n/s. These slugs could be detected within a reasonable time frame (200 s) when they were shielded by 10.16 cm of high-density polyethylene. The results obtained using the neutron noise analysis approach were compared with those obtained using the commonly used differential die-away analysis technique, a sensitive technique for detecting the presence of fissile materials by utilizing the prompt fission neutrons produced when the source neutrons from a neutron generator are completely diminished. For example, the time to detect 2 unshielded uranium slugs was 2.1 s when using the differential die-away analysis technique; it increased to 93 s for the neutron noise analysis approach. Although the noise analysis-based approach exhibits an overall performance which is not as good as that of differential die-away, neutron noise provides an alternative method for effective detection of special nuclear materials.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A proximity-focusing RICH detector for the ePIC Experiment at the EIC

The Electron-Proton/Ion Collider Experiment (ePIC) will be a large, multi-purpose detector to be installed at the Electron-Ion Collider (EIC) being built at Brookhaven National Laboratory. As robust particle identification (PID) capabilities are essential for fully realizing the EIC science program, ePIC contains several PID subsystems spanning different angular ranges. PID capability in the electron-going endcap is provided by a proximity-focusing Ring Imaging Cherenkov detector (pfRICH) designed to deliver at least 3σ separation between pions and kaons for momenta up to 7 GeV/ c . It will also aid with electron-hadron discrimination at low momentum and assist in the determination of the collision time (t 0 ). This contribution will summarize the design of the pfRICH as well as ongoing fabrication and component testing efforts. geant-based performance simulations will also be discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of a high voltage delivery system for noble liquid time projection chambers

Noble liquid time projection chambers (TPCs) are a leading technology in the detection of ionizing radiation, particularly in applications such as accelerator neutrino physics, dark matter detection, and neutrinoless double beta decay. This work addresses the design considerations for implementing stable high voltage (HV) systems within large noble liquid TPCs, with a focus on the nEXO experiment. Utilizing insights from prior HV research and experimental investigations, we outline factors influencing HV stability and discuss design choices to improve stability and prevent electrical discharges. A novel HV delivery system concept is presented, tailored for the nEXO TPC, which incorporates these design considerations while also meeting the stringent radiopurity requirements of the nEXO neutrinoless double beta decay search. These design considerations and their specific implementation towards a HV delivery system offer guidance to future experiments applying high voltage in noble liquid environments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗