DOE OSTI · 21124102
Diamond solid state ionization chambers for x-ray absorption spectroscopy applications
Abstract
The photoresponse of a diamond detector has been compared with a standard ionization chamber in x-ray absorption spectroscopy applications. A photoconductive device based on a nitrogen-doped single crystal diamond has been tested by synchrotron radiation. Time stability and linearity have been studied by x rays at 10 keV to assess its performances. Finally, extended x-ray absorption fine structure at the Fe K-edge was carried on a standard iron target using both the diamond device and the IC. Spectroscopical results have been compared including references to literature.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
De Sio, A, Bocci, A, Pace, E, Castellano, C, Cinque, G, Tartoni, N, D'Acapito, F. 2008-08-25. Diamond solid state ionization chambers for x-ray absorption spectroscopy applications. https://doi.org/10.1063/1.2976125. https://searchnasa.com/record/osti%3A21124102/diamond-solid-state-ionization-chambers-for-x-ray-absorption-spectroscopy-applications
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.