DOE OSTI · 22053949
Alignment effects on a neutron imaging system using coded apertures
Abstract
A high resolution neutron imaging system is being developed and tested on the OMEGA laser facility for inertial confinement fusion experiments. This diagnostic uses a coded imaging technique with a penumbral or an annular aperture. The sensitiveness of these techniques to misalignment was pointed out with both experiments and simulations. Results obtained during OMEGA shots are in good agreement with calculations performed with the Monte Carlo code GEANT4. Both techniques are sensitive to the relative position of the source in the field of view. The penumbral imaging technique then demonstrates to be less sensitive to misalignment compared to the ring. These results show the necessity to develop a neutron imaging diagnostic for megajoule class lasers taking into account our alignment capabilities on such facilities.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Thfoin, Isabelle, Landoas, Olivier, Caillaud, Tony, Vincent, Maxime, Bourgade, Jean-Luc, Rosse, Bertrand, Disdier, Laurent, Sangster, Thomas C, Glebov, Vladimir Yu, Pien, Greg, Armstrong, William. 2010-03-15. Alignment effects on a neutron imaging system using coded apertures. https://doi.org/10.1063/1.3331494
Cite the original work for its findings. Save a collection to share your selection of sources.