X-ray scatter correction method for dedicated breast computed tomography

Abstract

Purpose: To improve image quality and accuracy in dedicated breast computed tomography (BCT) by removing the x-ray scatter signal included in the BCT projections. Methods: The previously characterized magnitude and distribution of x-ray scatter in BCT results in both cupping artifacts and reduction of contrast and accuracy in the reconstructions. In this study, an image processing method is proposed that estimates and subtracts the low-frequency x-ray scatter signal included in each BCT projection postacquisition and prereconstruction. The estimation of this signal is performed using simple additional hardware, one additional BCT projection acquisition with negligible radiation dose, and simple image processing software algorithms. The high frequency quantum noise due to the scatter signal is reduced using a noise filter postreconstruction. The dosimetric consequences and validity of the assumptions of this algorithm were determined using Monte Carlo simulations. The feasibility of this method was determined by imaging a breast phantom on a BCT clinical prototype and comparing the corrected reconstructions to the unprocessed reconstructions and to reconstructions obtained from fan-beam acquisitions as a reference standard. One-dimensional profiles of the reconstructions and objective image quality metrics were used to determine the impact of the algorithm. Results: The proposed additional acquisition results in negligible additional radiation dose to the imaged breast ({approx}0.4% of the standard BCT acquisition). The processed phantom reconstruction showed substantially reduced cupping artifacts, increased contrast between adipose and glandular tissue equivalents, higher voxel value accuracy, and no discernible blurring of high frequency features. Conclusions: The proposed scatter correction method for dedicated breast CT is feasible and can result in highly improved image quality. Further optimization and testing, especially with patient images, is necessary to characterize its impact on clinical performance.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Sechopoulos, Ioannis. 2012-05-15. X-ray scatter correction method for dedicated breast computed tomography. https://doi.org/10.1118/1.4711749. https://searchnasa.com/record/osti%3A22098864/x-ray-scatter-correction-method-for-dedicated-breast-computed-tomography

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related reports

Cross sections of 147–149 Sm( 6 Li,x) reactions for the production of 149 Tb for targeted alpha therapy

Terbium-149g (t 1/2 = 4.12 h) is of particular interest for targeted alpha therapy cancer treatment due to its ability to decay via both alpha and positron emission, making it a potential theranostic nuclide. Due to many challenges facing its production, there are limited facilities worldwide that have demonstrated the ability to produce this nuclide in quantities sufficient for medical research. Since the Cyclotron Institute at Texas A&M University is a specialized accelerator facility capable of accelerating a wide variety of ions, we are investigating production pathway options. One of the major challenges facing its production is the known co-production of the excited isomeric state, 149m Tb (t 1/2 = 4.1 min). However, this state does not decay to the ground state of 149g Tb, negating any potential contribution to its yield. Due to its short-half life, the cross section for the population of this state has never been measured. After calculating several potential reaction yields using predictive models, the reactions of 147–149 Sm( 6 Li,xn) 149 Tb were identified as candidates. Lithium-6 beams of varied energies between 45-65 MeV were impinged on enriched 147 Sm, 148 Sm, and 149 Sm targets at the Cyclotron Institute at Texas A&M University, and the reaction products were measured immediately following irradiation using high-purity germanium detectors, enabling detection of both 149m Tb and 149g Tb. Cross sections for all nuclides produced in sufficient activity in these reactions were also measured and reported here. We conclude that the population of 149m Tb is much preferred over population of the ground state for these 6 Li-induced reactions, and it is necessary to explore other options for 149g Tb production.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Source function from two-particle correlation function through entropy-regularized Richardson-Lucy deblurring

Source functions are obtained from p – p and d – α correlation functions by applying the Richardson-Lucy (RL) deblurring to the Koonin-Pratt (KP) equation. To prevent fitting of noise in the correlation function, total-variation (TV) regularization is employed that has been effective in ordinary image restoration. TV alone cannot ensure normalization of the source functions. To ensure the latter, we propose a maximum-entropy regularized RL algorithm (MEM-RL). We outline the MEM-RL formalism and optimization strategy for the KP equation, demonstrating its effectiveness on both simulated and experimental data, including the p – p and d – α correlation functions.

62 RADIOLOGY AND NUCLEAR MEDICINE↗