Abstract
We demonstrate very high resolution photon spectroscopy with a microwave-multiplexed two-pixel transition-edge sensor (TES) array. We measured a {sup 153}Gd photon source and achieved an energy resolution of 63 eV full-width-at-half-maximum at 97 keV and an equivalent readout system noise of 86 pA/√(Hz) at the TES. The readout circuit consists of superconducting microwave resonators coupled to radio-frequency superconducting-quantum-interference-devices and transduces changes in input current to changes in phase of a microwave signal. We use flux-ramp modulation to linearize the response and evade low-frequency noise. This demonstration establishes one path for the readout of cryogenic X-ray and gamma-ray sensor arrays with more than 10{sup 3} elements and spectral resolving powers R=λ/Δλ>10{sup 3}.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Noroozian, Omid, Mates, John A. B., Bennett, Douglas A., Brevik, Justus A., Fowler, Joseph W., Gao, Jiansong, Hilton, Gene C., Horansky, Robert D., Irwin, Kent D., Schmidt, Daniel R., Vale, Leila R., Ullom, Joel N., Kang, Zhao. 2013-11-11. High-resolution gamma-ray spectroscopy with a microwave-multiplexed transition-edge sensor array. https://doi.org/10.1063/1.4829156. https://searchnasa.com/record/osti%3A22224085/high-resolution-gamma-ray-spectroscopy-with-a-microwave-multiplexed-transition-edge-sensor-array
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.