Deterministic absorbed dose estimation in computed tomography using a discrete ordinates method

Abstract

Purpose: Organ dose estimation for a patient undergoing computed tomography (CT) scanning is very important. Although Monte Carlo methods are considered gold-standard in patient dose estimation, the computation time required is formidable for routine clinical calculations. Here, the authors instigate a deterministic method for estimating an absorbed dose more efficiently. Methods: Compared with current Monte Carlo methods, a more efficient approach to estimating the absorbed dose is to solve the linear Boltzmann equation numerically. In this study, an axial CT scan was modeled with a software package, Denovo, which solved the linear Boltzmann equation using the discrete ordinates method. The CT scanning configuration included 16 x-ray source positions, beam collimators, flat filters, and bowtie filters. The phantom was the standard 32 cm CT dose index (CTDI) phantom. Four different Denovo simulations were performed with different simulation parameters, including the number of quadrature sets and the order of Legendre polynomial expansions. A Monte Carlo simulation was also performed for benchmarking the Denovo simulations. A quantitative comparison was made of the simulation results obtained by the Denovo and the Monte Carlo methods. Results: The difference in the simulation results of the discrete ordinates method and those of the Monte Carlo methods was found to be small, with a root-mean-square difference of around 2.4%. It was found that the discrete ordinates method, with a higher order of Legendre polynomial expansions, underestimated the absorbed dose near the center of the phantom (i.e., low dose region). Simulations of the quadrature set 8 and the first order of the Legendre polynomial expansions proved to be the most efficient computation method in the authors’ study. The single-thread computation time of the deterministic simulation of the quadrature set 8 and the first order of the Legendre polynomial expansions was 21 min on a personal computer. Conclusions: The simulation results showed that the deterministic method can be effectively used to estimate the absorbed dose in a CTDI phantom. The accuracy of the discrete ordinates method was close to that of a Monte Carlo simulation, and the primary benefit of the discrete ordinates method lies in its rapid computation speed. It is expected that further optimization of this method in routine clinical CT dose estimation will improve its accuracy and speed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Norris, Edward T., Liu, Xin, Hsieh, Jiang. 2015-07-15. Deterministic absorbed dose estimation in computed tomography using a discrete ordinates method. https://doi.org/10.1118/1.4922000

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Novel Method for Domestic Stable Isotope Production (CRADA #667) Final Report

This report presents the progress achieved on our chlorine isotope separation efforts in the CRADA #667 agreement. We used a microchannel distillation (MCD) packing structure to enrich HCl isotopes (H35Cl, H37Cl) and an isotachophoresis (ITP) separation to enrich ionic chloride species (35Cl- and 37Cl-). A 2 m tall MCD column was constructed and achieved 264 separation stages over several days, which is well above the goal of 50 stages outlined in the CRADA. The resulting height equivalent to a theoretical plate (HETP) was 0.76 cm. ITP tests were run ~ 24 hours and achieved a single-stage separation factor of 1.559.

07 ISOTOPE AND RADIATION SOURCES↗