Abstract
A grating groove density measuring system based on diffraction method has been developed to measure the two-dimensional groove density variations of the VLS gratings at NSRL. The measuring system can provide not only the basic data of the two-dimensional groove density variations for evaluating the imaging properties of grating, but also an important tool for improving the grating manufacturing technology. A commercially available concave grating with a curvature radius of 750 mm was measured to test the accuracy of the measuring system. Results show that the system has a measuring accuracy (ΔN/N) of about 5 × 10{sup −5}. Then the groove density variation parameters of the 1200 lines/mm VLS plane grating used in NSRL is measured by the system. The groove curvature of the VLS plane grating has been observed.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Du, Liangliang, Du, Xuewei, Wang, Qiuping. 2016-07-27. Two-dimensional groove density measurement for gratings by diffraction method. https://doi.org/10.1063/1.4952911