Purpose: Automate brachytherapy treatment plan quality check using Eclipse v13.6 scripting API based on pre-configured rules to minimize human error and maximize efficiency. Methods: The HDR Precheck system is developed based on a rules-driven approach using Eclipse scripting API. This system checks for critical plan parameters like channel length, first source position, source step size and channel mapping. The planned treatment time is verified independently based on analytical methods. For interstitial or SAVI APBI treatment plans, a Patterson-Parker system calculation is performed to verify the planned treatment time. For endobronchial treatments, an analytical formula from TG-59 is used. Acceptable tolerances were defined based on clinical experiences in our department. The system was designed to show PASS/FAIL status levels. Additional information, if necessary, is indicated appropriately in a separate comments field in the user interface. Results: The HDR Precheck system has been developed and tested to verify the treatment plan parameters that are routinely checked by the clinical physicist. The report also serves as a reminder or checklist for the planner to perform any additional critical checks such as applicator digitization or scenarios where the channel mapping was intentionally changed. It is expected to reduce the current manual plan check time from 15 minutes to <1 minute. Conclusion: Automating brachytherapy plan prechecks significantly reduces treatment plan precheck time and reduces human errors. When fully developed, this system will be able to perform TG-43 based second check of the treatment planning system’s dose calculation using random points in the target and critical structures. A histogram will be generated along with tabulated mean and standard deviation values for each structure. A knowledge database will also be developed for Brachyvision plans which will then be used for knowledge-based plan quality checks to further reduce treatment planning errors and increase confidence in the planned treatment.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Palaniswaamy, G, Morrow, A, Kim, S, Rangaraj, D. 2016-06-15. TU-D-201-06: HDR Plan Prechecks Using Eclipse Scripting API. https://doi.org/10.1118/1.4957472

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Deep Learning enabled spectral energy conversion for in situ exposure measurements

A detector-specific deep learning (DL) approach is presented for spectra-to-exposure conversion using large-format sodium iodide (NaI(Tl)) detectors deployed for in situ environmental radiation measurements in emergency response scenarios. Accurate determination of exposure from NaI spectra is challenging due to poor energy resolution, partial energy absorption, and the strong sensitivity of traditionally deployed analytical conversion methods to calibrated source geometry and pre-deployment assumptions. Here, to address these limitations, a multi-layer perceptron model was trained on a hybrid in situ /Monte Carlo dataset constructed to span a broad range of photon energies, spatial extents, and realistic deployment variability, representative of general in situ emergency response conditions. The DL model was evaluated against commonly fielded analytical approaches under matched simulation conditions, including a single-factor method, a G-function method, and a modeled pressurized ion chamber (PIC) baseline. This study was intentionally computational in scope to enable controlled, like-for-like comparisons between conversion techniques while minimizing confounding real-world variability. Comparison to the modeled PIC provides contextual benchmarking and is not intended as a field inter-comparison with deployed instruments. Across the evaluated 20 keV to 3 MeV energy range, the DL approach consistently exhibited higher accuracy and reduced variance relative to the analytical methods against a deterministically calculated exposure. This may indicate improved robustness to spectral complexity without reliance on source-, geometric-, or spectral region-specific optimization. While results do not represent real-world validation, the presented work demonstrates that deep learning may effectively learn the nonlinear detector response-to-exposure relationship for asymmetric NaI(Tl) detectors and offers a promising pathway for improving in situ exposure estimation using spectroscopic systems already integrated into initial real-time emergency response operations.

61 RADIATION PROTECTION AND DOSIMETRY↗

ACRRF High-Bay Dose Calculations using MCNP (Part B)

ACRR radiation outputs through vertical cavities have been documented in two reports. In Part B, the maximum dose from the unshielded central cavity, FREC-II, and NRS is calculated to inform the safety basis. The maximum dose is ≈1430 mrem per 300 MJ, or ≈40,930 rem/hr for full-power operation. A verification study completes the V&V of the modeling. Supplementary studies of variance-reduction techniques, model sensitivities, and aircraft dose above the ACRRF are included.

61 RADIATION PROTECTION AND DOSIMETRY↗

Logistics for the 2026 International Nuclear Accident Dosimetry Intercomparison Exercise

The purpose of this exercise is to test participants’ nuclear accident dosimeters (NADs) and personnel to the requirements established by American National Standards Institute (ANSI), Health Physics Society (HPS) standard ANSI/HPS-N13.3-2013 (R2025) Dosimetry for Criticality Accidents. IER-634 is the second nuclear accident dosimetry exercise taking place at the Armed Forces Radiobiology Research Institute (AFRRI) in Bethesda, MD, a Department of Defense (DOD) facility. This effort continues use of the high-dose neutron exposure test bed maintained by the DOD and represents the continued collaboration between the DOD and Department of Energy (DOE). Previous work at AFRRI included a reactor characterization (IER 484) in 2023 and a nuclear accident intercomparison exercise (IER-602) in 2024.

61 RADIATION PROTECTION AND DOSIMETRY↗