Connecting radiation-driven changes in structural, thermal, and mechanical properties in several medical device polymers

Abstract

Stemming from security threats and shortages in supply, a recent push has emerged to develop alternatives to radioisotope Cobalt-60 gamma radiation for sterilization of polymeric medical products, including electron beam (e-beam) and X-ray machine-based sources. However, before large-scale implementation, the effects of these non-isotope-based methods must be thoroughly investigated from several perspectives, involving their potential detrimental effects on the structural, thermal, and mechanical properties of polymers in medical devices. This paper investigates such effects in commonly used medical device polymers, including polypropylene homopolymer (PPH), polyolefin elastomer (POE), low-density polyethylene (LDPE), acrylonitrile butadiene styrene (ABS), and chlorobutyl rubber (CIIR). To connect radiation-driven changes in polymer structural and thermal properties and corresponding changes in mechanical properties, several characterization techniques were utilized, including gel permeation chromatography (GPC), Fourier transform infrared spectroscopy (FTIR), dynamic mechanical analysis (DMA) and differential scanning calorimetry (DSC). We found that the tensile strength of X-ray irradiated PPH, as well as X-ray, e-beam, and gamma irradiated CIIR decreases due to chain scission, which is corroborated by GPC and DMA. The GPC results of LDPE suggest chain scission, but interestingly without a corresponding decline in tensile strength. By comparison, POE and ABS appear to undergo further crosslinking with increasing irradiation dose, which generally strengthens them. Importantly, in the vast majority of the cases, there were small to no changes in properties upon changing radiation technologies (at a given dose), i.e., no radiation source dependence. However, in a few specific cases, radiation method dependencies did arise, for instance, in terms of the molecular weight of PPH being significantly increased upon X-ray irradiation, while the melting temperature, glass transition temperature, and elongation at break reduced compared to gamma. Overall, this study provides valuable insight into radiation-driven chemical and structural changes in polymers that produce changes in physical and functional properties.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hasan, Md Kamrul K., Staack, David, Pillai, Suresh D., Fifield, Leonard S., Pharr, Matt. 2024-01-28. Connecting radiation-driven changes in structural, thermal, and mechanical properties in several medical device polymers. https://doi.org/10.1016/j.polymdegradstab.2024.110677

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Cyclic moisture reactivation of calcium sorbents for long duration thermochemical energy storage

The transition to a flexible and reliable energy infrastructure, using electro-thermal energy generation technologies such as geothermal, concentrated solar power, and nuclear, usually demands simultaneous advancement of thermal energy storage (TES) to support on-demand electricity generation and industrial applications while mitigating the inherent intermittency of renewable energy sources and power outages from direct energy generation. Among TES technologies, thermochemical energy storage (TCES) based on calcium looping emerges as a compelling high-power energy storage candidate due to its high reaction enthalpy, compatibility with elevated operating temperatures, and abundance of low-cost materials. However, the long-term durability of calcium-based sorbents for TCES is hindered by surface sintering and particle aggregation, leading to performance degradation over repeated thermal cycles. This study explores a moisture hydration-based strategy to regenerate a degraded calcium sorbent and mitigate performance degradation for long duration TCES. The addition of moisture transforms calcium oxide into calcium hydroxide and produces intercalation water layers, associated with a regenerated surface area and reduced calcium oxide crystallite size. Both these effects are beneficial in restoring the sorbents' reactivity for carbonization. Additionally, an optimized hydration-assisted reactivation protocol balances the recovered energy storage capacity with heating penalty required for moisture removal from hydrated samples, resulting in an enhanced energy storage capacity up to 176% compared to benchmark sorbents that undergo cycling without reactivation after 60 cycles. In conclusion, these results highlight the potential of hydration-assisted reactivation to enhance the long-term performance of TCES, providing an effective pathway to advancing electro-thermal storage technologies.

36 MATERIALS SCIENCE↗