Miniaturized Magnetoelastic Sensor System

Abstract

This article describes the design, assembly, and implementation of a hand-held, magnetic-field-based sensor system that can be adapted for a variety of sensing applications. The miniaturized system is based on Chemical Identification by Magneto-Elastic Sensing (ChIMES) technology, which uses three concentric solenoid coils to wirelessly interrogate a sensor body comprised of a response material coupled to a magnetoelastic wire. The response material expands when it encounters a target, imposing mechanical stress on the wire and altering its magnetic permeability. The sensor bodies are passive, requiring no external power source, and they are small, measuring about 15 mm in length and 3.0 mm in diameter. Up to four sensor bodies can be configured as an evenly-spaced linear array. The sensor system operates by applying a low-frequency, current-stabilized, filtered triangle wave to a uniform-density excitation coil to switch the magnetic domains within the wire. Further, the responses from the sensors are picked up by a detection coil as stress-induced changes in the Faraday voltage, and the strong magnetic field induced by the excitation coil in the detection coil is nullified by a cancellation coil reverse-wound in series with the detection coil. The responses of the sensors in an array are separated in time by a linear gradient dc biasing coil. The sensors can be interrogated through metallic and nonmetallic barriers. The signals from the detection coil and the excitation coil are digitized by a pair of bipolar analog-to-digital converters (ADCs). A Raspberry Pi single-board computer (SBC) and associated software perform data acquisition and control all aspects of the sensor system hardware. The program allows the user to select the number of sensors in the array, the type of signal that is being collected, and the number of samples to take. The program also allows for signal processing of the sensor data, such as baseline correction. The program can differentiate sensor peaks from each other and calculate the magnitude of each sensor response with less than 1% error. The data are then displayed along with a graph of the signal.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Winslow, Joshua, Lamberti, Vincent E., Palmer, Michael P., Tokash, Justin C., DeVinney, Michael, Dallas, Tim. 2024-03-29. Miniaturized Magnetoelastic Sensor System. https://doi.org/10.1109/jsen.2024.3379880

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Performance evaluation of neutron noise analysis in detecting special nuclear materials

Reliable and rapid inspection techniques play a vital role in preventing illicit trafficking of special nuclear materials. Active interrogation systems using neutrons produced by portable, high-flux deuterium-deuterium or deuterium-tritium neutron generators are being actively developed as a secondary scanning tool for this purpose. In this study, a neutron noise analysis-based approach for detecting unshielded and shielded special nuclear materials by using a pulsed deuterium-tritium neutron generator was evaluated. Here, this approach analyzes the fluctuation of neutron counts. Its performance was quantified with regard to time-to-detection to achieve a minimum probability of detection of 99% and a probability of false alarm of less than 1% considering various amounts of special nuclear materials and different shielding configurations. It was demonstrated that this approach could detect 17 uranium slugs in 5 s given a neutron generator yield of 8.1 × 10 7 n/s. These slugs could be detected within a reasonable time frame (200 s) when they were shielded by 10.16 cm of high-density polyethylene. The results obtained using the neutron noise analysis approach were compared with those obtained using the commonly used differential die-away analysis technique, a sensitive technique for detecting the presence of fissile materials by utilizing the prompt fission neutrons produced when the source neutrons from a neutron generator are completely diminished. For example, the time to detect 2 unshielded uranium slugs was 2.1 s when using the differential die-away analysis technique; it increased to 93 s for the neutron noise analysis approach. Although the noise analysis-based approach exhibits an overall performance which is not as good as that of differential die-away, neutron noise provides an alternative method for effective detection of special nuclear materials.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A proximity-focusing RICH detector for the ePIC Experiment at the EIC

The Electron-Proton/Ion Collider Experiment (ePIC) will be a large, multi-purpose detector to be installed at the Electron-Ion Collider (EIC) being built at Brookhaven National Laboratory. As robust particle identification (PID) capabilities are essential for fully realizing the EIC science program, ePIC contains several PID subsystems spanning different angular ranges. PID capability in the electron-going endcap is provided by a proximity-focusing Ring Imaging Cherenkov detector (pfRICH) designed to deliver at least 3σ separation between pions and kaons for momenta up to 7 GeV/ c . It will also aid with electron-hadron discrimination at low momentum and assist in the determination of the collision time (t 0 ). This contribution will summarize the design of the pfRICH as well as ongoing fabrication and component testing efforts. geant-based performance simulations will also be discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of a high voltage delivery system for noble liquid time projection chambers

Noble liquid time projection chambers (TPCs) are a leading technology in the detection of ionizing radiation, particularly in applications such as accelerator neutrino physics, dark matter detection, and neutrinoless double beta decay. This work addresses the design considerations for implementing stable high voltage (HV) systems within large noble liquid TPCs, with a focus on the nEXO experiment. Utilizing insights from prior HV research and experimental investigations, we outline factors influencing HV stability and discuss design choices to improve stability and prevent electrical discharges. A novel HV delivery system concept is presented, tailored for the nEXO TPC, which incorporates these design considerations while also meeting the stringent radiopurity requirements of the nEXO neutrinoless double beta decay search. These design considerations and their specific implementation towards a HV delivery system offer guidance to future experiments applying high voltage in noble liquid environments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗