DOE OSTI · 1570285
Imaging effectiveness calculator for non-design microscope samples
Abstract
When attempting to integrate single-molecule fluorescence microscopy with microfabricated devices such as microfluidic channels, fabrication constraints may prevent using traditional coverslips. Instead, the fabricated devices may require imaging through material with a different thickness or index of refraction. Altering either can easily reduce the quality of the image formation (measured by the Strehl ratio) by a factor of 2 or more, reducing the signal-to-noise ratio accordingly. In such cases, successful detection of single-molecule fluorescence may prove difficult or impossible. Furthermore, we provide software to calculate the effect of non-design materials upon the Strehl ratio or ensquared energy and explore the impact of common materials used in microfabrication.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Anthony, Stephen M., Miller, Philip R., Timlin, Jerilyn A., Polsky, Ronen. 2019-07-26. Imaging effectiveness calculator for non-design microscope samples. https://doi.org/10.1364/ao.58.006027