Imaging effectiveness calculator for non-design microscope samples

Abstract

When attempting to integrate single-molecule fluorescence microscopy with microfabricated devices such as microfluidic channels, fabrication constraints may prevent using traditional coverslips. Instead, the fabricated devices may require imaging through material with a different thickness or index of refraction. Altering either can easily reduce the quality of the image formation (measured by the Strehl ratio) by a factor of 2 or more, reducing the signal-to-noise ratio accordingly. In such cases, successful detection of single-molecule fluorescence may prove difficult or impossible. Furthermore, we provide software to calculate the effect of non-design materials upon the Strehl ratio or ensquared energy and explore the impact of common materials used in microfabrication.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Anthony, Stephen M., Miller, Philip R., Timlin, Jerilyn A., Polsky, Ronen. 2019-07-26. Imaging effectiveness calculator for non-design microscope samples. https://doi.org/10.1364/ao.58.006027

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗