DOE OSTI · 1603648
A versatile Johansson-type tender x-ray emission spectrometer
Abstract
We present a high energy resolution x-ray spectrometer for the tender x-ray regime (1.6–5.0 keV) that was designed and operated at Stanford Synchrotron Radiation Lightsource. The instrument is developed on a Rowland geometry (500 mm of radius) using cylindrically bent Johansson analyzers and a position sensitive detector. By placing the sample inside the Rowland circle, the spectrometer operates in an energy-dispersive mode with a subnatural line-width energy resolution (~0.32 eV at 2400 eV), even when an extended incident x-ray beam is used across a wide range of diffraction angles (~30° to 65°). The spectrometer is enclosed in a vacuum chamber, and a sample chamber with independent ambient conditions is introduced to enable a versatile and fast-access sample environment (e.g., solid/gas/liquid samples, in situ cells, and radioactive materials). The design, capabilities, and performance are presented and discussed.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Nowak, S. H., Armenta, R., Schwartz, C. P., Gallo, A., Abraham, B., Garcia-Esparza, A. T., Biasin, E., Prado, A., Maciel, A., Zhang, D., Day, D., Christensen, S., Kroll, T., Alonso-Mori, R., Nordlund, D., Weng, T. -C., Sokaras, D.. 2020-03-03. A versatile Johansson-type tender x-ray emission spectrometer. https://doi.org/10.1063/1.5121853
Cite the original work for its findings. Save a collection to share your selection of sources.