Interlaboratory Study of Digital Volume Correlation Error Due to X-Ray Computed Tomography Equipment and Scan Parameters: an Update from the DVC Challenge

Abstract

Background: The quality of Digital Volume Correlation (DVC) full-field displacement measurements depends directly on the characteristics of the X-ray Computed Tomography (XCT) equipment, and scan procedures used to acquire the tomographic images. Objective: In this work, we seek to experimentally study the effects of XCT equipment and tomographic scan procedures on the quality of these images for DVC analysis, and to survey the level of DVC error that may be achieved using standard XCT operating procedures. Methods: Six participants in an interlaboratory study acquired high-quality XCT scans of a syntactic foam before and after rigid body motion. The resulting images were correlated using commercial DVC software to quantify error sources due to random image noise, reconstruction artifacts, as well as systematic spatial or temporal distortion. Results: In the absence of rigid body motion, the standard deviation of the displacement measurements ranged from 0.012 to 0.043 voxels using a moderate subvolume size, indicating that subvoxel measurement resolution could readily be achieved with a variety of XCT equipment and scan recipes. Comparison of consecutive scans without rigid body motion showed transient dilatational displacement gradients due to self-heating of the X-ray source and/or thermal expansion of the foam. Evaluation of the scans after rigid body motion showed significant, machine-specific spatial distortion in the displacement fields of up to 0.5 voxels; new approaches to remove this error need to be developed. Conclusions: Analysis of the scan protocols used in the interlaboratory study, as well as a complementary parametric sensitivity study, showed that the DVC error was strongly influenced by the XCT equipment, but could be mitigated by adjusting the total scan duration.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Croom, Brendan P., Burden, Diana, Jin, Helena, Vonk, N. H., Hoefnagels, J. P. M., Smaniotto, B., Hild, F., Quintana, E., Sun, Q., Nie, X., Li, Xiaodong. 2020-09-09. Interlaboratory Study of Digital Volume Correlation Error Due to X-Ray Computed Tomography Equipment and Scan Parameters: an Update from the DVC Challenge. https://doi.org/10.1007/s11340-020-00653-x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗