DOE OSTI · 1909806
Method for calculating radon activity and radon rejection using a beta-gamma detector
Abstract
Radon interferes with concentration measurements used by atmospheric radioxenon systems. We demonstrate a method to quantify the amount of radon that is present in the detectors, the impact of radon activity on the minimum-detectable-concentrations, and how to determine the needed radon rejection levels. An example calculation shows a radon rejection level of 10 5 is sufficient to limit impact on the detector sensitivity. We anticipate this method will give analysist a better understanding of radon present in their measurements and allow system designers to tailor their systems’ radon rejection better for its location.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Mayer, Michael F., Cooper, Matthew W., Ely, James H., Hayes, James C.. 2022-08-29. Method for calculating radon activity and radon rejection using a beta-gamma detector. https://doi.org/10.1007/s10967-022-08480-1
Cite the original work for its findings. Save a collection to share your selection of sources.