DOE OSTI · 22062377
Time-resolved soft x-ray absorption setup using multi-bunch operation modes at synchrotrons
Abstract
Here, we report on a novel experimental apparatus for performing time-resolved soft x-ray absorption spectroscopy in the sub-ns time scale using non-hybrid multi-bunch mode synchrotron radiation. The present setup is based on a variable repetition rate Ti:sapphire laser (pump pulse) synchronized with the {approx}500 MHz x-ray synchrotron radiation bunches and on a detection system that discriminates and singles out the significant x-ray photon pulses by means of a custom made photon counting unit. The whole setup has been validated by measuring the time evolution of the L{sub 3} absorption edge during the melting and the solidification of a Ge single crystal irradiated by an intense ultrafast laser pulse. These results pave the way for performing synchrotron time-resolved experiments in the sub-ns time domain with variable repetition rate exploiting the full flux of the synchrotron radiation.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Stebel, L, Sigalotti, P, Ressel, B, Cautero, G, Malvestuto, M, Capogrosso, V, Bondino, F, Magnano, E, Parmigiani, F. 2011-12-15. Time-resolved soft x-ray absorption setup using multi-bunch operation modes at synchrotrons. https://doi.org/10.1063/1.3669787
Cite the original work for its findings. Save a collection to share your selection of sources.