Search NASA⌕ Search

DOE OSTI · 22098667

Reference dosimetry for helical tomotherapy: Practical implementation and a multicenter validation

Abstract

Purpose: The aim of this study was to implement a protocol for reference dosimetry in tomotherapy and to validate the beam output measurements with an independent dosimetry system. Methods: Beam output was measured at the reference depth of 10 cm in water for the following three cases: (1) a 5 x 10 cm{sup 2} static machine specific reference field (MSR), (2) a rotational 5 x 10 cm{sup 2} field without modulation and no tabletop in the beam, (3) a plan class specific reference (PCSR) field defined as a rotational homogeneous dose delivery to a cylindrical shaped target volume: plan with modulation and table-top movement. The formalism for reference dosimetry of small and nonstandard fields [Med.Phys.35: 5179-5186, 2008] and QA recommendations [Med.Phys.37: 4817-4853, 2010] were adopted in the dose measurement protocol. All ionization chamber measurements were verified independently using alanine/EPR dosimetry. As a pilot study, the beam output was measured on tomotherapy Hi-art systems at three other centers and directly compared to the centers specifications and to alanine dosimetry. Results: For the four centers, the mean static output at a depth of 10 cm in water and SAD = 85 cm, measured with an A1SL chamber following the TG-148 report was 6.238 Gy/min {+-} 0.058 (1 SD); the rotational output was 6.255 Gy/min {+-} 0.069 (1 SD). The dose stated by the center was found in good agreement with the measurements of the visiting team: D{sub center}/D{sub visit} = 1.000 {+-} 0.003 (1 SD). The A1SL chamber measurements were all in good agreement with Alanine/EPR dosimetry. Going from the static reference field to the rotational/non modulated field the dose rate remains constant within 0.2% except for one center where a deviation of 1.3% was detected. Conclusions: Following the TG-148 report, beam output measurements in water at the reference depth using a local protocol, as developed at different centers, was verified. The measurements were found in good agreement with alanine/EPR dosimetry. The presented methodology may provide a good concept for reference dosimetry.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

De Ost, B., Schaeken, B., Vynckier, S., Sterpin, E., Van den Weyngaert, D.. 2011-11-15. Reference dosimetry for helical tomotherapy: Practical implementation and a multicenter validation. https://doi.org/10.1118/1.3651496

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Deep Learning enabled spectral energy conversion for in situ exposure measurements

A detector-specific deep learning (DL) approach is presented for spectra-to-exposure conversion using large-format sodium iodide (NaI(Tl)) detectors deployed for in situ environmental radiation measurements in emergency response scenarios. Accurate determination of exposure from NaI spectra is challenging due to poor energy resolution, partial energy absorption, and the strong sensitivity of traditionally deployed analytical conversion methods to calibrated source geometry and pre-deployment assumptions. Here, to address these limitations, a multi-layer perceptron model was trained on a hybrid in situ /Monte Carlo dataset constructed to span a broad range of photon energies, spatial extents, and realistic deployment variability, representative of general in situ emergency response conditions. The DL model was evaluated against commonly fielded analytical approaches under matched simulation conditions, including a single-factor method, a G-function method, and a modeled pressurized ion chamber (PIC) baseline. This study was intentionally computational in scope to enable controlled, like-for-like comparisons between conversion techniques while minimizing confounding real-world variability. Comparison to the modeled PIC provides contextual benchmarking and is not intended as a field inter-comparison with deployed instruments. Across the evaluated 20 keV to 3 MeV energy range, the DL approach consistently exhibited higher accuracy and reduced variance relative to the analytical methods against a deterministically calculated exposure. This may indicate improved robustness to spectral complexity without reliance on source-, geometric-, or spectral region-specific optimization. While results do not represent real-world validation, the presented work demonstrates that deep learning may effectively learn the nonlinear detector response-to-exposure relationship for asymmetric NaI(Tl) detectors and offers a promising pathway for improving in situ exposure estimation using spectroscopic systems already integrated into initial real-time emergency response operations.

61 RADIATION PROTECTION AND DOSIMETRY↗

ACRRF High-Bay Dose Calculations using MCNP (Part B)

ACRR radiation outputs through vertical cavities have been documented in two reports. In Part B, the maximum dose from the unshielded central cavity, FREC-II, and NRS is calculated to inform the safety basis. The maximum dose is ≈1430 mrem per 300 MJ, or ≈40,930 rem/hr for full-power operation. A verification study completes the V&V of the modeling. Supplementary studies of variance-reduction techniques, model sensitivities, and aircraft dose above the ACRRF are included.

61 RADIATION PROTECTION AND DOSIMETRY↗

Logistics for the 2026 International Nuclear Accident Dosimetry Intercomparison Exercise

The purpose of this exercise is to test participants’ nuclear accident dosimeters (NADs) and personnel to the requirements established by American National Standards Institute (ANSI), Health Physics Society (HPS) standard ANSI/HPS-N13.3-2013 (R2025) Dosimetry for Criticality Accidents. IER-634 is the second nuclear accident dosimetry exercise taking place at the Armed Forces Radiobiology Research Institute (AFRRI) in Bethesda, MD, a Department of Defense (DOD) facility. This effort continues use of the high-dose neutron exposure test bed maintained by the DOD and represents the continued collaboration between the DOD and Department of Energy (DOE). Previous work at AFRRI included a reactor characterization (IER 484) in 2023 and a nuclear accident intercomparison exercise (IER-602) in 2024.

61 RADIATION PROTECTION AND DOSIMETRY↗