Automatic CT simulation optimization for radiation therapy: A general strategy

Abstract

Purpose: In radiation therapy, x-ray computed tomography (CT) simulation protocol specifications should be driven by the treatment planning requirements in lieu of duplicating diagnostic CT screening protocols. The purpose of this study was to develop a general strategy that allows for automatically, prospectively, and objectively determining the optimal patient-specific CT simulation protocols based on radiation-therapy goals, namely, maintenance of contouring quality and integrity while minimizing patient CT simulation dose. Methods: The authors proposed a general prediction strategy that provides automatic optimal CT simulation protocol selection as a function of patient size and treatment planning task. The optimal protocol is the one that delivers the minimum dose required to provide a CT simulation scan that yields accurate contours. Accurate treatment plans depend on accurate contours in order to conform the dose to actual tumor and normal organ positions. An image quality index, defined to characterize how simulation scan quality affects contour delineation, was developed and used to benchmark the contouring accuracy and treatment plan quality within the predication strategy. A clinical workflow was developed to select the optimal CT simulation protocols incorporating patient size, target delineation, and radiation dose efficiency. An experimental study using an anthropomorphic pelvis phantom with added-bolus layers was used to demonstrate how the proposed prediction strategy could be implemented and how the optimal CT simulation protocols could be selected for prostate cancer patients based on patient size and treatment planning task. Clinical IMRT prostate treatment plans for seven CT scans with varied image quality indices were separately optimized and compared to verify the trace of target and organ dosimetry coverage. Results: Based on the phantom study, the optimal image quality index for accurate manual prostate contouring was 4.4. The optimal tube potentials for patient sizes of 38, 43, 48, 53, and 58 cm were 120, 140, 140, 140, and 140 kVp, respectively, and the corresponding minimum CTDIvol for achieving the optimal image quality index 4.4 were 9.8, 32.2, 100.9, 241.4, and 274.1 mGy, respectively. For patients with lateral sizes of 43–58 cm, 120-kVp scan protocols yielded up to 165% greater radiation dose relative to 140-kVp protocols, and 140-kVp protocols always yielded a greater image quality index compared to the same dose-level 120-kVp protocols. The trace of target and organ dosimetry coverage and the γ passing rates of seven IMRT dose distribution pairs indicated the feasibility of the proposed image quality index for the predication strategy. Conclusions: A general strategy to predict the optimal CT simulation protocols in a flexible and quantitative way was developed that takes into account patient size, treatment planning task, and radiation dose. The experimental study indicated that the optimal CT simulation protocol and the corresponding radiation dose varied significantly for different patient sizes, contouring accuracy, and radiation treatment planning tasks.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Li, Hua, Chen, Hsin-Chen, Tan, Jun, Gay, Hiram, Michalski, Jeff M., Mutic, Sasa, Yu, Lifeng, Anastasio, Mark A., Low, Daniel A.. 2014-03-15. Automatic CT simulation optimization for radiation therapy: A general strategy. https://doi.org/10.1118/1.4866377

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Cross sections of 147–149 Sm( 6 Li,x) reactions for the production of 149 Tb for targeted alpha therapy

Terbium-149g (t 1/2 = 4.12 h) is of particular interest for targeted alpha therapy cancer treatment due to its ability to decay via both alpha and positron emission, making it a potential theranostic nuclide. Due to many challenges facing its production, there are limited facilities worldwide that have demonstrated the ability to produce this nuclide in quantities sufficient for medical research. Since the Cyclotron Institute at Texas A&M University is a specialized accelerator facility capable of accelerating a wide variety of ions, we are investigating production pathway options. One of the major challenges facing its production is the known co-production of the excited isomeric state, 149m Tb (t 1/2 = 4.1 min). However, this state does not decay to the ground state of 149g Tb, negating any potential contribution to its yield. Due to its short-half life, the cross section for the population of this state has never been measured. After calculating several potential reaction yields using predictive models, the reactions of 147–149 Sm( 6 Li,xn) 149 Tb were identified as candidates. Lithium-6 beams of varied energies between 45-65 MeV were impinged on enriched 147 Sm, 148 Sm, and 149 Sm targets at the Cyclotron Institute at Texas A&M University, and the reaction products were measured immediately following irradiation using high-purity germanium detectors, enabling detection of both 149m Tb and 149g Tb. Cross sections for all nuclides produced in sufficient activity in these reactions were also measured and reported here. We conclude that the population of 149m Tb is much preferred over population of the ground state for these 6 Li-induced reactions, and it is necessary to explore other options for 149g Tb production.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Source function from two-particle correlation function through entropy-regularized Richardson-Lucy deblurring

Source functions are obtained from p – p and d – α correlation functions by applying the Richardson-Lucy (RL) deblurring to the Koonin-Pratt (KP) equation. To prevent fitting of noise in the correlation function, total-variation (TV) regularization is employed that has been effective in ordinary image restoration. TV alone cannot ensure normalization of the source functions. To ensure the latter, we propose a maximum-entropy regularized RL algorithm (MEM-RL). We outline the MEM-RL formalism and optimization strategy for the KP equation, demonstrating its effectiveness on both simulated and experimental data, including the p – p and d – α correlation functions.

62 RADIOLOGY AND NUCLEAR MEDICINE↗